An investigation of the relationship between ultrasound echo enhancement and Doppler frequency shift using a pulsatile arterial flow phantom.

J Petrick1, M Zomack, R Schlief.   

Abstract

RATIONALE AND
OBJECTIVES: Based on the echo-enhancing effect of microbubbles, various agents have been developed to improve the diagnostic confidence in patients with inadequate Doppler signals. In the clinical trials of the echo enhancer Levovist, there were a few isolated cases in which the maximum flow velocities of the enhanced Doppler spectra appeared to higher than the velocities in the nonenhanced baseline spectra. This raised the concern that echo enhancement could give a false indication of maximum flow velocity. This study investigated whether a systematic association between echo enhancement and velocity shift exists.
METHODS: A pulsatile flow phantom that simulated the elasticity of blood vessels and the acoustic attenuation of extravascular tissue was used to compare enhanced with unenhanced Doppler spectra under accurately reproducible flow conditions. The experiments were carried out with varied sound attenuation and evaluated with dedicated spectral analysis software that included a special averaging tool.
RESULTS: Levovist enhanced the Doppler signal by 16 to 31 dB, and the enhanced and unenhanced power spectra presented identical distribution of spectral power density under all flow conditions.
CONCLUSIONS: The measurements gave no evidence that echo enhancement with Levovist falsifies the Doppler measurement of flow velocity.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9101358     DOI: 10.1097/00004424-199704000-00006

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  2 in total

1.  Effect of microbubble contrast on intracranial blood flow velocity assessed by transcranial Doppler.

Authors:  Nicola Logallo; Annette Fromm; Ulrike Waje-Andreassen; Lars Thomassen; Knut Matre
Journal:  J Ultrasound       Date:  2014-01-29

2.  The effect of contrast material on transcranial Doppler evaluation of normal middle cerebral artery peak systolic velocity.

Authors:  H G Khan; P Gailloud; R O Bude; J B Martin; K T Szopinski; C Khaw; D A Rüfenacht; K J Murphy
Journal:  AJNR Am J Neuroradiol       Date:  2000-02       Impact factor: 3.825

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.