Literature DB >> 9077430

Comparison of the sensitivity of four rapid assays for the detection of antibodies to HIV-1/HIV-2 during seroconversion.

H H Samdal1, B G Gutigard, D Labay, S I Wiik, K Skaug, A G Skar.   

Abstract

OBJECTIVES: To compare the sensitivity of four rapid assays for the detection of antibodies against HIV-1 during early seroconversion.
METHODS: Four rapid assays for the detection of antibodies to HIV-1/HIV-2 (SUDS HIV 1 + 2, TestPack HIV-1/HIV-2, HIV-SPOT and CAPILLUS HIV-1/HIV-2) were evaluated on 38 sera derived from 14 HIV-1 seroconverters, formerly tested with Abbott second- and third-generation HIV-1/HIV-2 enzyme immunoassay (EIA) and Diagnostic Biotechnology HIV blot 2.2. EIA-negative sera had also been investigated by HIV-1 antigen testing and polymerase chain reaction (PCR) analysis detecting HIV-1 proviral DNA.
RESULTS: On 16 sera which were Abbott second-generation EIA negative, the SUDS assay rated highest with six positive results. Four sera were TestPack positive. No specimens in this group were HIV-SPOT positive or CAPILLUS positive. Of the remaining 22 sera (Abbott second- and third-generation EIA positive), all but two were positive in all tests.
CONCLUSIONS: The study revealed differences in the sensitivities of the rapid assays in the early phase of seroconversion. The most sensitive assay, SUDS, was even more sensitive than the Abbott third-generation EIA, while TestPack and the EIA were equally sensitive. Based on these findings, we suggest the inclusion of these assays in the supplemental testing for detection of antibodies to HIV.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 9077430     DOI: 10.1016/s0928-0197(96)00244-9

Source DB:  PubMed          Journal:  Clin Diagn Virol        ISSN: 0928-0197


  8 in total

1.  HIV nucleic acid amplification testing versus rapid testing: it is worth the wait. Testing preferences of men who have sex with men.

Authors:  Joshua D O'Neal; Matthew R Golden; Bernard M Branson; Joanne D Stekler
Journal:  J Acquir Immune Defic Syndr       Date:  2012-08-01       Impact factor: 3.731

2.  A simple whole-blood test for detecting antibodies to human immunodeficiency virus.

Authors:  W Schramm; S E Wade; G Barriga Angulo; P Castillo Torres; A Burgess-Cassler
Journal:  Clin Diagn Lab Immunol       Date:  1998-03

3.  Evaluation of SAMBA II: A Qualitative and Semiquantitative HIV Point-of-Care Nucleic Acid Test.

Authors:  Lauren R Violette; Andy Cornelius-Hudson; Madison Snidarich; Lisa A Niemann; Sonny Michael Assennato; Allyson Ritchie; Neha Goel; Pollyanna R Chavez; Steven F Ethridge; David A Katz; Helen Lee; Kevin P Delaney; Joanne D Stekler
Journal:  J Acquir Immune Defic Syndr       Date:  2022-04-15       Impact factor: 3.771

4.  HIV testing in a high-incidence population: is antibody testing alone good enough?

Authors:  Joanne D Stekler; Paul D Swenson; Robert W Coombs; Joan Dragavon; Katherine K Thomas; Catherine A Brennan; Sushil G Devare; Robert W Wood; Matthew R Golden
Journal:  Clin Infect Dis       Date:  2009-08-01       Impact factor: 9.079

5.  Assessment of rapid tests for detection of human immunodeficiency virus-specific antibodies in recently infected individuals.

Authors:  Brian Louie; Ernest Wong; Jeffrey D Klausner; Sally Liska; Frederick Hecht; Terri Dowling; Martha Obeso; Susan S Phillips; Mark W Pandori
Journal:  J Clin Microbiol       Date:  2008-01-30       Impact factor: 5.948

6.  Utility of rapid antibody tests to exclude HIV-1 infection among infants and children aged <18 months in a low-resource setting.

Authors:  Ann M Buchanan; Behzad Nadjm; Ben Amos; George Mtove; David Sifuna; Coleen K Cunningham; John A Crump; Hugh Reyburn
Journal:  J Clin Virol       Date:  2012-08-24       Impact factor: 3.168

7.  Validation, performance under field conditions, and cost-effectiveness of Capillus HIV-1/HIV-2 and determine HIV-1/2 rapid human immunodeficiency virus antibody assays using sequential and parallel testing algorithms in Tanzania.

Authors:  Meghan K Mayhood; Isaac A Afwamba; Christopher O Odhiambo; Epimack Ndanu; Nathan M Thielman; Anne B Morrissey; John F Shao; Brian Wells Pence; John A Crump
Journal:  J Clin Microbiol       Date:  2008-10-15       Impact factor: 5.948

8.  Relative accuracy of serum, whole blood, and oral fluid HIV tests among Seattle men who have sex with men.

Authors:  Joanne D Stekler; Joshua D O'Neal; Aric Lane; Fred Swanson; Janine Maenza; Claire E Stevens; Robert W Coombs; Joan A Dragavon; Paul D Swenson; Matthew R Golden; Bernard M Branson
Journal:  J Clin Virol       Date:  2013-12       Impact factor: 3.168