Literature DB >> 8968857

Body composition during weight loss in obese patients estimated by dual energy X-ray absorptiometry and by total body potassium.

H W Hendel1, A Gotfredsen, T Andersen, L Højgaard, J Hilsted.   

Abstract

OBJECTIVE: To validate the ability of DXA to measure total body composition before and after weight loss and the composition of the lost body mass.
DESIGN: Cross sectional and follow-up study of body composition before and after a weight loss of 10.6 +/- 6.8 kg.
SUBJECTS: 31 obese subjects with a mean body weight of 105.2 +/- 15.2 kg at baseline, and a mean body weight of 94.6 +/- 16.5 kg at follow-up. MEASUREMENTS: Body composition was measured by dual X-ray absorptiometry, total body potassium counting, and high precision scales before and after a weight loss.
RESULTS: DXA and the scales showed a strong linear relation (r = 1). At baseline, however, DXA underestimated the body weight by a maximum of 3.2 kg because the subjects were too large for the scan table. After weight loss body weight measurements were accurate. Before and after weight loss the linear relations between DXA and TBK for FFM were strong (r = 0.92 and 0.93). Bland and Altman plots showed limits of agreement of +/-9 kg before and after weight loss; DXA underestimated FFM in women and overestimated FFM in men. DXA accounted for 80% of the lost body weight. The composition of the lost body mass did not differ from that estimated by TBK (7.6% FFM and 92.4% FM by TBK; 11% FFM and 89% FM by DXA).
CONCLUSION: DXA estimates accurately the body composition and the composition of weight loss in groups of obese subjects. However, the scan table may be too small for patients weighing more than 95 kg.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8968857

Source DB:  PubMed          Journal:  Int J Obes Relat Metab Disord


  6 in total

Review 1.  The effects of weight loss approaches on bone mineral density in adults: a systematic review and meta-analysis of randomized controlled trials.

Authors:  S Soltani; G R Hunter; A Kazemi; S Shab-Bidar
Journal:  Osteoporos Int       Date:  2016-05-06       Impact factor: 4.507

2.  Comparison of bioimpedance spectroscopy and dual energy X-ray absorptiometry for assessing body composition changes in obese children during weight loss.

Authors:  Eline Vermeiren; Marijke Ysebaert; Kim Van Hoorenbeeck; Luc Bruyndonckx; Kristof Van Dessel; Maria Van Helvoirt; Ann De Guchtenaere; Benedicte De Winter; Stijn Verhulst; Annelies Van Eyck
Journal:  Eur J Clin Nutr       Date:  2020-09-11       Impact factor: 4.016

3.  Relationship between body composition and bone mineral density in women with and without osteoporosis: relative contribution of lean and fat mass.

Authors:  Saverio Gnudi; Emanuela Sitta; Nicoletta Fiumi
Journal:  J Bone Miner Metab       Date:  2007-08-25       Impact factor: 2.626

4.  Measurement of body composition changes with weight loss in postmenopausal women: comparison of methods.

Authors:  A K Mahon; M G Flynn; H B Iglay; L K Stewart; C A Johnson; B K McFarlin; W W Campbell
Journal:  J Nutr Health Aging       Date:  2007 May-Jun       Impact factor: 4.075

5.  Accuracy of body composition measurements by dual energy x-ray absorptiometry in underweight patients with chronic intestinal disease and in lean subjects.

Authors:  Kent Valentin Haderslev; Pernille Heldager Haderslev; Michael Staun
Journal:  Dyn Med       Date:  2005-01-04

6.  Measuring body composition in overweight individuals by dual energy x-ray absorptiometry.

Authors:  Rhonda A Brownbill; Jasminka Z Ilich
Journal:  BMC Med Imaging       Date:  2005-03-04       Impact factor: 1.930

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.