E Garcia-Cimbrelo1, J M Martinez-Sayanes, A Minuesa, L Munuera.
Abstract
The clinical status of 83 Mittelmeier ceramic-ceramic (Mark II) cementless total hip prostheses (Autophor, Osteo AG, Selzach, Switzerland) implanted between 1978 and 1984 was analyzed. Retrieved tissue from the revised hips was studied histologically. The mean patient age was 47.5 years (range, 19-71 years). One or both components of 12 hips were revised (16% at 16 years in the survival study). Eleven acetabular components and seven stems were revised, with three of the sockets fractured. The mean follow-up period for the patients with unrevised hips was 12.3 years. Osseointegration was uncommon in both the cup and stem. Radiographic cup loosening was observed in 53% of hips at 16 years in the survival study, with most having a complete radiolucent line of 2 mm around the cup. Radiographic stem loosening appeared in 15% of hips at 16 years. Fibrous stable fixation was present in 64% of stems. Significant radiographic acetabular wear was not observed in any hip. The Mittelmeier Mark II prosthesis had worse long-term results than reported with cemented or other uncemented prostheses. Although there was no significant acetabular wear, intracellular ceramic wear debris was seen in the retrieved tissues.Entities:
Mesh:
Year: 1996 PMID: 8934316 DOI: 10.1016/s0883-5403(96)80176-8
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757