The effects of therapeutic contact lenses on intraocular pressure measurement.

G D Scibilia1, W H Ehlers, P C Donshik.   

Abstract

PURPOSE: To evaluate the accuracy of intraocular pressure (IOP) measurements taken through three commonly used therapeutic contact lenses, using both the Tono-Pen and pneumatonometer.
METHODS: We measured the IOP of ten normal eyes and ten eyes with abnormal anterior segments using Tono-Pen, pneumatonometry, and Goldman application tonometry. Three different soft contact lenses (Acuvue, 04, and Permalens) were sequentially placed on the eyes and repeat measurements were taken. The readings obtained for each contact lens were compared to the baseline readings taken without a lens in place in order to assess the effect that each contact lens had on the accuracy of IOP measurements.
RESULTS: All three contact lenses permitted accurate IOP measurements in eyes with both normal and abnormal anterior segments. Variations in water content and central thickness did not affect the accuracy of IOP measurements.
CONCLUSIONS: Both Tono-Pen and pneumatonometry proved to be equally reliable in recording accurate IOP measurements through therapeutic contact lenses.

Entities:  

Mesh:

Year:  1996        PMID: 8906384

Source DB:  PubMed          Journal:  CLAO J        ISSN: 0733-8902


  7 in total

Review 1.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

2.  Intraocular Pressure After 2 Hours of Small-Diameter Scleral Lens Wear.

Authors:  Cherie B Nau; Muriel M Schornack; Jay W McLaren; Arthur J Sit
Journal:  Eye Contact Lens       Date:  2016-11       Impact factor: 2.018

3.  The effects of the modulus of the lens material on intraocular pressure measurement through soft contact lenses.

Authors:  S Boyraz; I Güngör
Journal:  Ir J Med Sci       Date:  2012-12-04       Impact factor: 1.568

4.  Comparison of intraocular pressure measurement with Scheimpflug-based noncontact tonometer with and without hydrogel contact lenses.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Debarun Dutta; Maneck D Nicolsan; Sriharsha Nagaraj; Rajesh S Kumar
Journal:  Indian J Ophthalmol       Date:  2015-04       Impact factor: 1.848

5.  Ocular surface displacement with and without contact lenses during non-contact tonometry.

Authors:  Ulfah Rimayanti; Yoshiaki Kiuchi; Shohei Uemura; Joji Takenaka; Hideki Mochizuki; Makoto Kaneko
Journal:  PLoS One       Date:  2014-04-29       Impact factor: 3.240

6.  Detecting IOP Fluctuations in Glaucoma Patients.

Authors:  Brenda Nuyen; Kaweh Mansouri
Journal:  Open Ophthalmol J       Date:  2016-02-29

7.  The measurement of intraocular pressure over positive soft contact lenses by rebound tonometry.

Authors:  Fabrizio Zeri; Mario De Cusatis; Luigi Lupelli; Peter Graham Swann
Journal:  J Optom       Date:  2015-10-21
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.