Literature DB >> 8885660

Metal-on-metal articulation for artificial hip joints: laboratory study and clinical results.

R M Streicher1, M Semlitsch, R Schön, H Weber, C Rieker.   

Abstract

As wear is inevitable with artificial joint replacement, it has to be minimized to avoid possible aseptic loosening following osteolysis due to particle-initiated foreign body reaction. Co-Cr-Mo-C alloys have a long history with only minimum wear when articulating with themselves. This investigation shows that the choice of the alloy has an effect on the wear rate of this articulation couple. Tribological studies in a screening device, a pendulum apparatus and a hip joint simulator showed a marked influence of the environment as well as the diameter of the implants with metal-on-metal articulation. A wear-resistant combination with low friction characteristics has been developed by using a wrought Co-Cr-Mo-C alloy and reducing the implant diameter to 28 mm. Clinical wear rates are comparable with laboratory data and demonstrate the potential of the metal-on-metal articulation to solve the problem of wear-induced osteolysis of hip joint endoprostheses.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8885660     DOI: 10.1243/PIME_PROC_1996_210_416_02

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  9 in total

1.  Quantitative analysis of the wear and wear debris from low and high carbon content cobalt chrome alloys used in metal on metal total hip replacements.

Authors:  J L Tipper; P J Firkins; E Ingham; J Fisher; M H Stone; R Farrar
Journal:  J Mater Sci Mater Med       Date:  1999-06       Impact factor: 3.896

2.  Cup loosening after cemented Metasul® total hip replacement: a retrieval analysis.

Authors:  Christophe Nich; Moussa Hamadouche
Journal:  Int Orthop       Date:  2010-06-11       Impact factor: 3.075

3.  Sporting and physical activity following hip resurfacing.

Authors:  Noel E Fisher; Vijay V Killampalli; Rakesh K Kundra; Nikolas A Jagodzinski; Keshav Mathur; Alex D Reading
Journal:  Int Orthop       Date:  2010-05-30       Impact factor: 3.075

4.  Delta ceramic-on-alumina ceramic articulation in primary THA: prospective, randomized FDA-IDE study and retrieval analysis.

Authors:  Adolph V Lombardi; Keith R Berend; Brian E Seng; Ian C Clarke; Joanne B Adams
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

5.  Wear testing of a DJOA finger prosthesis in vitro.

Authors:  Thomas J Joyce
Journal:  J Mater Sci Mater Med       Date:  2010-02-11       Impact factor: 3.896

6.  Cardiac cobaltism: a rare complication after bilateral metal-on-metal total hip arthroplasty.

Authors:  John R Martin; Luke Spencer-Gardner; Christopher L Camp; John M Stulak; Rafael J Sierra
Journal:  Arthroplast Today       Date:  2015-11-21

7.  Tribology of total hip arthroplasty prostheses: What an orthopaedic surgeon should know.

Authors:  Claude B Rieker
Journal:  EFORT Open Rev       Date:  2017-03-13

8.  High-tech hip implant for wireless temperature measurements in vivo.

Authors:  Georg Bergmann; Friedmar Graichen; Jörn Dymke; Antonius Rohlmann; Georg N Duda; Philipp Damm
Journal:  PLoS One       Date:  2012-08-22       Impact factor: 3.240

Review 9.  Imaging Metal-on-Metal Hip Replacements: the Norwich Experience.

Authors:  Marianna S Thomas; James A Wimhurst; John F Nolan; Andoni P Toms
Journal:  HSS J       Date:  2013-08-24
  9 in total