Literature DB >> 8883648

Anatomic evaluation of the circle of Willis: MR angiography versus intraarterial digital subtraction angiography.

K W Stock1, S Wetzel, E Kirsch, G Bongartz, W Steinbrich, E W Radue.   

Abstract

PURPOSE: To evaluate the reliability of source images and maximum intensity projection images of MR angiography in showing the arterial segments of the circle of Willis.
METHODS: In 62 patients, 526 arterial segments of the circle of Willis were determined to be present, partially present, or absent by blinded observers evaluating MR angiographic source images and maximum intensity projection images. Vessel diameter was measured on source images. These results were then compared with the results from intraarterial digital subtraction angiography.
RESULTS: MR angiographic maximum intensity projection images had a sensitivity of 87% and a specificity of 88% and MR angiographic source images had a sensitivity of 89% and a specificity of 63% in depicting the presence of a vessel segment. The positive predictive value of an arterial segment with a diameter of at least 1 mm was 99%.
CONCLUSION: MR angiography is a sensitive technique for detecting the anatomy of the circle of Willis. Maximum intensity projection images are more specific than source images. An arterial segment with a diameter of at least 1 mm on the source image is almost always present and patent.

Entities:  

Mesh:

Year:  1996        PMID: 8883648      PMCID: PMC8338732     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  26 in total

1.  MR angiography in patients with subarachnoid hemorrhage: adequate to evaluate vasospasm-induced vascular narrowing?

Authors:  Elke Hattingen; Stella Blasel; Richard Dumesnil; Hartmut Vatter; Friedhelm E Zanella; Stefan Weidauer
Journal:  Neurosurg Rev       Date:  2010-06-08       Impact factor: 3.042

2.  Assessment of the collateral function of the circle of Willis: three-dimensional time-of-flight MR angiography compared with transcranial color-coded duplex sonography.

Authors:  Arjan W J Hoksbergen; Charles B L Majoie; Frans-Jan H Hulsmans; Dink A Legemate
Journal:  AJNR Am J Neuroradiol       Date:  2003-03       Impact factor: 3.825

3.  The predictive value of 3D time-of-flight MR angiography in assessment of brain arteriovenous malformation obliteration after radiosurgery.

Authors:  D R Buis; J C J Bot; F Barkhof; D L Knol; F J Lagerwaard; B J Slotman; W P Vandertop; R van den Berg
Journal:  AJNR Am J Neuroradiol       Date:  2011-11-17       Impact factor: 3.825

4.  Quantification and assessment of extracranial and intracranial occlusive disease.

Authors:  Muhammad S Hussain; Rishi Gupta
Journal:  J Vasc Interv Neurol       Date:  2008-01

5.  Quality-evaluation scheme for cerebral time-resolved 3D contrast-enhanced MR angiography techniques.

Authors:  H Raoult; J-C Ferré; X Morandi; B Carsin-Nicol; M Carsin; M Cuggia; M Law; J-Y Gauvrit
Journal:  AJNR Am J Neuroradiol       Date:  2010-05-06       Impact factor: 3.825

6.  Morphology of the cerebral arterial circle in the prenatal and postnatal period of Serbian population.

Authors:  Ljiljana Vasović; Milena Trandafilović; Ivan Jovanović; Slađana Ugrenović; Slobodan Vlajković; Miroslav Milić; Gordana Đorđević
Journal:  Childs Nerv Syst       Date:  2013-05-24       Impact factor: 1.475

7.  Mechanisms of bihemispheric brain infarctions in the anterior circulation on diffusion-weighted images.

Authors:  Kozue Saito; Hiroshi Moriwaki; Hiroshi Oe; Kotaro Miyashita; Kazuyuki Nagatsuka; Satoshi Ueno; Hiroaki Naritomi
Journal:  AJNR Am J Neuroradiol       Date:  2005-04       Impact factor: 3.825

8.  Comparison Between 3D TOF Magnetic Resonance Angiography and Intraarterial Digital Subtraction Angiography in Imaging the Circle of Willis.

Authors:  Rozita Mohd Ghazali; Ibrahim Lutfi Shuaib
Journal:  Malays J Med Sci       Date:  2003-01

9.  Optimal MRI sequence for identifying occlusion location in acute stroke: which value of time-resolved contrast-enhanced MRA?

Authors:  A Le Bras; H Raoult; J-C Ferré; T Ronzière; J-Y Gauvrit
Journal:  AJNR Am J Neuroradiol       Date:  2015-03-12       Impact factor: 3.825

10.  Accuracy of pre- and postcontrast 3D time-of-flight MR angiography in patients with acute ischemic stroke: correlation with catheter angiography.

Authors:  H Ishimaru; M Ochi; M Morikawa; H Takahata; Y Matsuoka; T Koshiishi; T Fujimoto; A Egawa; K Mitarai; T Murakami; M Uetani
Journal:  AJNR Am J Neuroradiol       Date:  2007-05       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.