Accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients.

M M Hirschl1, M Binder, H Herkner, A Bur, M Brunner, D Seidler, H G Stühlinger, A N Laggner.   

Abstract

OBJECTIVE: To evaluate the accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients.
DESIGN: Prospective data collection.
SETTING: Emergency department in a 2,000-bed hospital. PATIENTS: Thirty-nine patients admitted to the emergency department requiring invasive arterial blood pressure monitoring were enrolled to the study protocol.
INTERVENTIONS: Continuous noninvasive blood pressure measurement was performed on the middle phalanx of the second and third finger, using a test instrument which provides continuous arterial waveform display with the use of a finger cuff. Invasive mean arterial blood pressure measurement was done by cannulation of the radial artery and direct transduction of the systemic arterial pressure waveform.
MEASUREMENTS AND MAIN RESULTS: Three thousand one hundred eighteen pairs of simultaneous finger cuff and intra-arterial blood pressure measurements were collected in 1-min intervals from 39 patients over a total of 51.8 hrs. The overall discrepancy between both measurements was 0.10 mm Hg. The standard deviation of the differences was +/- 5.02 mm Hg. The mean bias in patients treated with catecholamines was 0.01 mm Hg and was not different from the bias observed in patients without catecholamines (mean bias: 0.23 mm Hg; p > .22). Whereas 95% of all comparisons between finger cuff and intra-arterial measurement had a discrepancy < or = +/- 10 mm Hg, 4.7% had a discrepancy between +/- 10.1 to 15 mm Hg and 0.3% exhibited a discrepancy > +/- 15 mm Hg. In 29 (74%) patients, the duration of errors > 10 mm Hg was < or = 1 min. In seven (18%) patients, the duration of errors > 10 mm Hg was between 2 to 3 mins and in three (8%) patients, the errors lasted for > 3 mins.
CONCLUSIONS: Our data provide a guide to the accuracy and reliability of noninvasive finger blood pressure measurements in critically ill patients. Although most test instrument measurements were reliable, in 8% of all patients large discrepancies (> 10 mm Hg) between both measurements with a duration of > 3 mins were noted. Concerning the considerable risk for arterial cannulation, our preliminary data demonstrate that the test instrument (PORTAPRES, TNO Biomedical Instrumentation Research Unit; The Netherlands) is an advance in noninvasive monitoring of critically ill patients and may be useful in most emergency clinical settings.

Entities:  

Mesh:

Year:  1996        PMID: 8874306     DOI: 10.1097/00003246-199610000-00014

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  6 in total

1.  The mean prehospital machine; accurate prehospital non-invasive blood pressure measurement in the critically ill patient.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2010-06-08       Impact factor: 2.502

2.  Validation of arterial blood pressures observed from the patient monitor; a tool for prehospital research.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2009-12-18       Impact factor: 2.502

3.  Inter-day reliability of blood pressure and cerebral blood flow velocities in persons with spinal cord injury and intact controls.

Authors:  Jill M Wecht; Joseph P Weir; William A Bauman
Journal:  J Spinal Cord Med       Date:  2016-02-17       Impact factor: 1.985

4.  The mean machine; accurate non-invasive blood pressure measurement in the critically ill patient.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2009-08-12       Impact factor: 2.502

5.  Positive airway pressure improves nocturnal beat-to-beat blood pressure surges in obesity hypoventilation syndrome with obstructive sleep apnea.

Authors:  Jason R Carter; Ida T Fonkoue; Daniela Grimaldi; Leila Emami; David Gozal; Colin E Sullivan; Babak Mokhlesi
Journal:  Am J Physiol Regul Integr Comp Physiol       Date:  2016-01-27       Impact factor: 3.619

6.  Increasing sleep duration to lower beat-to-beat blood pressure: a pilot study.

Authors:  Monika Haack; Jorge Serrador; Daniel Cohen; Norah Simpson; Hans Meier-Ewert; Janet M Mullington
Journal:  J Sleep Res       Date:  2012-11-22       Impact factor: 3.981

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.