Perioperative on-site haemoglobin determination: as accurate as laboratory values?

M Jaeger1, T Ashbury, M Adams, P Duncan.   

Abstract

PURPOSE: This study assessed the accuracy of photometer based haemoglobin (Hb) determination technology (HemoCue) when used by different anaesthetists in situations of rapidly changing Hb values during anaesthesia.
METHODS: (Part 1) In the laboratory, repeated measurements were done on 16 split samples of blood using both the Hematology Analyzer (CELL-DYN 3500 System, Abbot Laboratories, San Jose, California) and the photometer. (Part 2) Twelve patients had blood samples drawn from an arterial line for simultaneous Hb determination in the hospital laboratory and by the photometer. At the same time, capillary samples were taken from the patient's earlobe for Hb determination by the photometer. All sample collection and photometer measurements were done by the same operator. (Part 3) The Part 2 protocol was then repeated with different anaesthetists performing both the sampling and the photometer measurements. Statistical comparison was by ANOVA and a two-tailed paired t-test.
RESULTS: (Part 1) Samples determined by the photometer and the laboratory were highly correlated (r2 = 1.0, P < 0.001). The average error of each method was similar ( < 4%). (Part 2) Using a 2-tailed paired t-test, the photometer arterial measurements were not different from the laboratory measurements, however the photometer capillary measurements were consistently approximately 8% higher (P = 0.003). (Part 3) When multiple operators performed the sampling there were no differences on arterial or capillary samples (r2 = 0.942, r2 = 0.851 respectively), although the variance was greater.
CONCLUSIONS: The HemoCue haemoglobinometer has sufficient accuracy to support treatment decisions regarding blood transfusions.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8840058     DOI: 10.1007/BF03013031

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  4 in total

1.  A clinical evaluation of the Hemocue haemoglobinometer using capillary, venous and arterial samples.

Authors:  P P Chen; T G Short; D H Leung; T E Oh
Journal:  Anaesth Intensive Care       Date:  1992-11       Impact factor: 1.669

2.  An evaluation of the HemoCue for correcting the haemoglobin value of lipaemic samples.

Authors:  P D Nicholls
Journal:  Med Lab Sci       Date:  1990-07

3.  Evaluation of portable haemoglobinometer in general practice.

Authors:  R G Neville
Journal:  Br Med J (Clin Res Ed)       Date:  1987-05-16

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

  4 in total
  4 in total

1.  Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room.

Authors:  Jan Zatloukal; Jiri Pouska; Jakub Kletecka; Richard Pradl; Jan Benes
Journal:  J Clin Monit Comput       Date:  2015-10-27       Impact factor: 2.502

2.  Reliability of HemoCue in patients with gastrointestinal bleeding.

Authors:  Andry Van de Louw; Nadine Lasserre; François Drouhin; Stéphane Thierry; Lucien Lecuyer; Daniel Caen; Alain Tenaillon
Journal:  Intensive Care Med       Date:  2006-12-05       Impact factor: 17.440

Review 3.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

4.  Feasibility of performing multiple point of care testing for HIV anti-retroviral treatment initiation and monitoring from multiple or single fingersticks.

Authors:  Natasha Gous; Lesley Scott; Joachim Potgieter; Lumka Ntabeni; Sharon Enslin; Ronel Newman; Wendy Stevens
Journal:  PLoS One       Date:  2013-12-20       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.