Measured scattered x-ray energy spectra for simulated irradiation geometries in diagnostic radiology.

N W Marshall1, K Faulkner, H Warren.   

Abstract

In this study, scattered x-ray distributions were produced by irradiating an anthropomorphic pelvis phantom under fluoroscopic conditions using incident beams generated at tube potentials between 61 kVp and 112 kVp. Both overcouch and undercouch x-ray tube orientations were used when irradiating the phantom. The energy spectrum of the scattered x rays was measured with a germanium x-ray detector. Two measurement geometries were employed: (i) the detector placed at 90 degrees to the incident x-ray beam to determine the energy spectra incident on the trunk region of staff and (ii) the detector placed at 45 degrees to the x-ray beam, measuring spectra incident on the head and neck region. The effect of irradiation area on the scattered spectra was also investigated. Spectral distributions, along with mean energy and half-value layer (HVL) in mm Al, are presented for each spectrum. Energy spectra measured at 90 degrees to the incident x-ray beam were found to have HVLs approximately 10%-15% greater than the corresponding primary incident spectrum, for both overcouch and undercouch irradiations. The magnitude of the irradiation area had negligible effect on the mean energy and HVL of the spectra.

Entities:  

Mesh:

Year:  1996        PMID: 8839423     DOI: 10.1118/1.597690

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Investigation of scatter fractions for estimating leakage dose in medical X-ray imaging facilities.

Authors:  Kimiya Noto; Kichirou Koshida; Hiroji Iida; Tomoyuki Yamamoto; Ikuo Kobayashi; Chikako Kawabata
Journal:  Radiol Phys Technol       Date:  2009-04-08

2.  High filtration in interventional practices reduces patient radiation doses but not always scatter radiation doses.

Authors:  Roberto M Sanchez; Eliseo Vano; Pablo Salinas; Nieves Gonzalo; Javier Escaned; Jose M Fernández
Journal:  Br J Radiol       Date:  2020-11-24       Impact factor: 3.039

3.  Organ-specific external dose coefficients and protective apron transmission factors for historical dose reconstruction for medical personnel.

Authors:  Steven L Simon
Journal:  Health Phys       Date:  2011-07       Impact factor: 2.922

4.  Evaluation of stray radiation to the operator for five hand-held dental X-ray devices.

Authors:  Richard Smith; Richard Tremblay; Graeme M Wardlaw
Journal:  Dentomaxillofac Radiol       Date:  2019-02-12       Impact factor: 2.419

5.  Characterisation and mapping of scattered radiation fields in interventional radiology theatres.

Authors:  M Nowak; P Carbonez; M Krauss; F R Verdun; J Damet
Journal:  Sci Rep       Date:  2020-10-30       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.