End-tidal PCO2 monitoring via nasal cannulae in pediatric patients: accuracy and sources of error.

R H Friesen1, M Alswang.   

Abstract

OBJECTIVE: To assess the correlation and accuracy of end-tidal PCO2 (PetCO2) sampled via nasal cannulae in pediatric patients by comparison to the criterion standard PaCO2, and to identify sources of error during PetCO2 monitoring via nasal cannulae.
METHODS: PetCO2 was monitored continuously by sampling end-tidal gas through nasal cannulae that had been designed and manufactured for this purpose in spontaneously breathing children undergoing conscious or deep sedation during either cardiac catheterization (n = 43) or critical care (n = 54). When both the capnographic wave form and the PetCO2 value had been stable for at least 10 minutes, the PetCO2 value was recorded while blood was drawn from an indwelling arterial line for PaCO2 measurement. The effects of age, weight, respiratory rate, oxygen delivery system, airway obstruction, mouth breathing, and cyanotic heart disease were evaluated by linear regression analysis and calculation of absolute bias (PaCO2-PetCO2).
RESULTS: Mouth breathing, airway obstruction, oxygen delivery through the ipsilateral nasal cannula, and cyanotic heart disease adversely affected accuracy. In patients without those factors, PetCO2 correlated well with PaCO2 (R2 = 0.994), and absolute bias was 3.0 +/- 1.8 mmHg.
CONCLUSIONS: Several factors-some controllable and all recognizable-affect the accuracy of PetCO2 monitored via nasal cannulae in pediatric patients. When these factors are not present, PetCO2 correlates well with PaCO2 and appears to be a useful monitor of ventilatory status during conscious or deep sedation.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8823636     DOI: 10.1007/bf02078136

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  19 in total

1.  End tidal carbon dioxide and respiratory rate measurement during conscious sedation through a nasal cannula.

Authors:  W E Ackerman; J C Phero; D Reaume
Journal:  Anesth Prog       Date:  1990-07

2.  Estimation of arterial carbon dioxide by end-tidal and transcutaneous PCO2 measurements in ventilated children.

Authors:  Y Sivan; M K Eldadah; T E Cheah; C J Newth
Journal:  Pediatr Pulmonol       Date:  1992-03

3.  Conscious sedation in the emergency department: the value of capnography and pulse oximetry.

Authors:  S W Wright
Journal:  Ann Emerg Med       Date:  1992-05       Impact factor: 5.721

4.  Relationship between arterial carbon dioxide and end-tidal carbon dioxide when a nasal sampling port is used.

Authors:  S E McNulty; J Roy; M Torjman; J L Seltzer
Journal:  J Clin Monit       Date:  1990-04

5.  Can the Raman scattering analyzer compete with mass spectrometers: an affirmative reply.

Authors:  D R Westenskow; D L Coleman
Journal:  J Clin Monit       Date:  1989-01

6.  A simple, easy, and inexpensive method for monitoring ETCO2 through nasal cannulae.

Authors:  J M Goldman
Journal:  Anesthesiology       Date:  1987-10       Impact factor: 7.892

7.  Factors influencing capnography in the Bain circuit.

Authors:  N Gravenstein; S Lampotang; J E Beneken
Journal:  J Clin Monit       Date:  1985-01

8.  Changes in carbon dioxide tension and oxygen saturation during deep sedation for paediatric cardiac catheterization.

Authors:  R H Friesen; M Alswang
Journal:  Paediatr Anaesth       Date:  1996       Impact factor: 2.556

9.  End-tidal PCO2 measurements sampled at the distal and proximal ends of the endotracheal tube in infants and children.

Authors:  J M Badgwell; M E McLeod; J Lerman; R E Creighton
Journal:  Anesth Analg       Date:  1987-10       Impact factor: 5.108

10.  Accuracy of expiratory carbon dioxide measurements using the coaxial and circle breathing circuits in small subjects.

Authors:  R A Schieber; A Namnoum; A Sugden; A L Saville; R A Orr
Journal:  J Clin Monit       Date:  1985-07
View more
  9 in total

Review 1.  Critical review of non-invasive respiratory monitoring in medical care.

Authors:  M Folke; L Cernerud; M Ekström; B Hök
Journal:  Med Biol Eng Comput       Date:  2003-07       Impact factor: 2.602

Review 2.  Capnometry in the prehospital setting: are we using its potential?

Authors:  Dejan Kupnik; Pavel Skok
Journal:  Emerg Med J       Date:  2007-09       Impact factor: 2.740

3.  Clinical evaluation of a novel respiratory rate monitor.

Authors:  Peter J Lee
Journal:  J Clin Monit Comput       Date:  2015-04-22       Impact factor: 2.502

4.  Assessment of noninvasive acoustic respiration rate monitoring in patients admitted to an Emergency Department for drug or alcoholic poisoning.

Authors:  Youcef Guechi; Amélie Pichot; Denis Frasca; Fatima Rayeh-Pelardy; Jean-Yves Lardeur; Olivier Mimoz
Journal:  J Clin Monit Comput       Date:  2015-01-23       Impact factor: 2.502

5.  Comparative provocation test of respiratory monitoring methods.

Authors:  Mia Folke; Fredrik Granstedt; Bertil Hök; Håkan Scheer
Journal:  J Clin Monit Comput       Date:  2002-02       Impact factor: 2.502

6.  Risk factors for bradypnea in a historical cohort of surgical patients receiving fentanyl-based intravenous analgesia.

Authors:  Takashi Shiota; Hideaki Kawanishi; Satoki Inoue; Junji Egawa; Masahiko Kawaguchi
Journal:  JA Clin Rep       Date:  2018-06-13

7.  Acoustic respiration rate and pulse oximetry-derived respiration rate: a clinical comparison study.

Authors:  Michal E Eisenberg; Dalia Givony; Raz Levin
Journal:  J Clin Monit Comput       Date:  2018-11-26       Impact factor: 2.502

8.  Advanced Monitoring Is Associated with Fewer Alarm Events During Planned Moderate Procedure-Related Sedation: A 2-Part Pilot Trial.

Authors:  Richard L Applegate; John Lenart; Mathew Malkin; Minhthy N Meineke; Silvana Qoshlli; Monica Neumann; J Paul Jacobson; Alison Kruger; Jeffrey Ching; Mohammad Hassanian; Michael Um
Journal:  Anesth Analg       Date:  2016-04       Impact factor: 5.108

9.  Anesthesia Management for Pediatrics with Congenital Heart Diseases Who Undergo Cardiac Catheterization in China.

Authors:  Chun-Mei Xie; Yun-Tai Yao
Journal:  J Interv Cardiol       Date:  2021-01-30       Impact factor: 2.279

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.