Calculation of computed tomography dose index to effective dose conversion factors based on measurement of the dose profile along the fan shaped beam.

J T Jansen1, J Geleijns, D Zweers, F W Schultz, J Zoetelief.   

Abstract

The variation in computed tomography dose index (CTDI) to effective dose conversion factors between different types of CT scanner is large (i.e. a factor of about 2 due to differences in beam shaping filters). Consequently, scanner specific conversion factors have to be applied. For some types of scanner, however, detailed information on the construction of beam shaping filters is not provided by the manufacturers. It is of interest to investigate the use of measured dose profiles for the calculation of conversion factors. Based upon measured dose profiles, two appropriate photon spectra selected on the basis of measured half value layers, gender specific adult phantoms Adam and Eva, and the Monte Carlo neutron and photon radiation transport code (MCNP), organ and effective dose conversion factors are calculated. To validate the method, a comparison is made between results for measured and calculated beam profiles for a Philips Tomoscan 350. The results in terms of effective dose per slice per unit of CTDI are compared with published data. Relative difference in conversion factors per slice averaged over all slices used for the calculations is 13 +/- 4% between the two spectra, 10.2 +/- 0.2% between measured and calculated beam profiles and 50 +/- 191% between the phantoms of different gender. The relative difference between the averaged results for the Adam and Eva phantoms and published results for a hermaphrodite phantom is on average equal to or less than 15 +/- 13%, depending on the spectrum and beam profile used, although larger differences can occur for specific slices. It is concluded that CTDI to effective dose conversion factors can be derived on the basis of measured beam profiles.

Entities:  

Mesh:

Year:  1996        PMID: 8785619     DOI: 10.1259/0007-1285-69-817-33

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  3 in total

1.  Low-dose CT colonography in children: initial experience, technical feasibility, and utility.

Authors:  Sudha Anupindi; James Perumpillichira; Diego Jaramillo; Michael E Zalis; Esther J Israel
Journal:  Pediatr Radiol       Date:  2005-03-24

2.  Low-dose multidetector-row CT angiography of the infra-renal aorta and lower extremity vessels: image quality and diagnostic accuracy in comparison with standard DSA.

Authors:  Francesco Fraioli; Carlo Catalano; Alessandro Napoli; Marco Francone; Fiammetta Venditti; Massimiliano Danti; Federica Pediconi; Roberto Passariello
Journal:  Eur Radiol       Date:  2005-06-30       Impact factor: 5.315

3.  Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner.

Authors:  J Geleijns; M Salvadó Artells; P W de Bruin; R Matter; Y Muramatsu; M F McNitt-Gray
Journal:  Phys Med Biol       Date:  2009-05-06       Impact factor: 3.609

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.