Clinical comparison of the Keeler Pulsair 2000, American Optical MkII and Goldmann applanation tonometers.

S W Mackie1, J L Jay, R Ackerley, G Walsh.   

Abstract

The aim of our study was to evaluate the performance of both the Keeler Pulsair 2000 and the American Optical (AO) MkII non-contact tonometers (NCT) and compare these to the reference Goldmann standard using the same group of patients. Forty-five patients (89 eyes) receiving medical treatment for primary open angle glaucoma had their intraocular pressure (IOP) measured with each instrument in a random order using five experienced observers. In the IOP range of the sample (6-27 mmHg) the difference between means for each tonometer was small. The Pulsair 2000 (mean = 19.06 mmHg, SD 6.28) read slightly higher than Goldmann (mean = 18.01 mmHg, SD 4.88), whereas the AO MkII read slightly lower (mean = 16.27 mmHg, SD 5.93). However, all differences were statistically significant (P < 0.001, repeat measures ANOVA). Correlation coefficients for Pulsair 2000 versus Goldmann was r = 0.82 and for AO MkII versus Goldmann was r = 0.85. In addition, the repeatability for each NCT was assessed using 10 consecutive measurements on a further 10 subjects. The Pulsair 2000 showed significantly greater variation of IOP (mean SD 3.43 mmHg) than the AO MkII (mean SD 1.76 mmHg), confirming the need for taking at least four readings per eye with the Pulsair 2000. These variations are greater than that generally observed with Goldmann, and possible explanations why this occurs are discussed. In conclusion, subject to these limitations, both NCTs should be useful for measuring IOP as part of a screening protocol for glaucoma.

Entities:  

Mesh:

Year:  1996        PMID: 8762780

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  6 in total

1.  A comparison of four methods of tonometry: method agreement and interobserver variability.

Authors:  P-A Tonnu; T Ho; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

2.  Effects of corneal thickness on intraocular pressure measured with three different tonometers.

Authors:  Hiroki Murase; Akira Sawada; Kiyofumi Mochizuki; Tetsuya Yamamoto
Journal:  Jpn J Ophthalmol       Date:  2009-01-30       Impact factor: 2.447

3.  On the precision and reliability of IOP measurements.

Authors:  J M Gilchrist
Journal:  Br J Ophthalmol       Date:  1996-07       Impact factor: 4.638

4.  Dynamic contour tonometry over silicone hydrogel contact lens.

Authors:  Andrew K C Lam; Jimmy S H Tse
Journal:  J Optom       Date:  2013-08-28

5.  Clinical comparison of Pulsair non-contact tonometer and Goldmann applanation tonometer in Indian population.

Authors:  Shalini Mohan; Satyaprakash Tiwari; Arvind Jain; Jaya Gupta; Surendra Kumar Sachan
Journal:  J Optom       Date:  2013-07-26

6.  The applicability of correction factor for corneal thickness on non-contact tonometer measured intraocular pressure in LASIK treated eyes.

Authors:  Jitendra Jethani; Paaraj Dave; Monica Jethani; Yogesh Desai; Purvi Patel
Journal:  Saudi J Ophthalmol       Date:  2015-11-19
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.