Early postoperative rise in defibrillation threshold in patients with nonthoracotomy defibrillation lead systems: attenuation with biphasic shock waveforms.

D Schwartzman1, D J Callans, C D Gottlieb, J Heo, F E Marchlinski.   

Abstract

INTRODUCTION: In patients with non-thoracotomy defibrillation lead (NTL) systems coupled with monophasic shock waveforms, the defibrillation threshold (DFT) rises early after implantation. There is little information regarding features predictive of the DFT rise, or DFT changes early after implantation of NTL systems coupled with biphasic shock waveforms. METHODS AND
RESULTS: DFT measurements were performed serially at implantation, prior to hospital discharge (mean 4 +/- 3 days), and at follow-up (mean 49 +/- 22 days) in 146 patients with an NTL system. Factors were assessed for association with a "clinically important" early postimplantation DFT rise, defined as a rise of > or = 2 energy steps (2 to 4 J per step; > or = 5 J total). A clinically important early postimplantation DFT rise occurred in 48 patients (33%). Univariate predictors of the rise included the monophasic shock waveform, the Medtronic Transvene lead system, the presence of a subcutaneous defibrillation patch, and the number of shocks delivered during the implantation procedure. However, the only independent predictor of a clinically important DFT rise was the monophasic shock waveform (F = 18, P < 0.001). For the monophasic patient group (n = 79), the incidence of a DFT rise was 53% (n = 42). For the biphasic patient group (n = 67), the incidence of a DFT rise was 9% (n = 6). The clinical characteristics of the monophasic and biphasic groups were not significantly different, nor were their DFTs at implantation. Among a subgroup of 18 consecutive patients who underwent serial DFT testing utilizing both monophasic and biphasic waveforms, the incidence of a clinically important DFT rise with monophasic (n = 9,50%) was higher than with biphasic shocks (n = 3,17%; P = 0.05).
CONCLUSION: NTL systems coupled with biphasic shock waveforms have an attenuated incidence of a clinically important DFT rise early after implantation, relative to patients with NTL systems coupled to monophasic waveforms.

Entities:  

Mesh:

Year:  1996        PMID: 8743754     DOI: 10.1111/j.1540-8167.1996.tb00555.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  5 in total

1.  Submuscular versus subcutaneous pectoral implantation of cardioverter-defibrillators: effect on high voltage pathway impedance and defibrillation efficacy.

Authors:  D Iskos; K Lock; K G Lurie; G J Fahy; S Petersen-Stejskal; D G Benditt
Journal:  J Interv Card Electrophysiol       Date:  1998-03       Impact factor: 1.900

2.  Early postoperative rise in defibrillation threshold associated with hematoma formation with unipolar defibrillation system.

Authors:  D Y Yazmajian; D Schwartzman; D J Callans; C D Gottieb; F E Marchlinski
Journal:  J Interv Card Electrophysiol       Date:  1997-09       Impact factor: 1.900

3.  Effect of general anesthesia on the defibrillation energy requirement in patients undergoing defibrillator implantation.

Authors:  B P Knight; F Pelosi; M Flemming; F Morady; S A Strickberger
Journal:  J Interv Card Electrophysiol       Date:  1999-12       Impact factor: 1.900

4.  [Influence of waveform and configuration of electrodes on the defibrillation threshold of implantable cardioverter-defibrillators].

Authors:  M Block; D Hammel; G Breithardt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

5.  A randomized comparison of defibrillation thresholds in the right ventricular outflow tract versus right ventricular apex.

Authors:  Michael Mollerus; Margaret Lipinski; Thomas Munger
Journal:  J Interv Card Electrophysiol       Date:  2008-05-10       Impact factor: 1.900

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.