Effect of general anesthesia on the defibrillation energy requirement in patients undergoing defibrillator implantation.

B P Knight1, F Pelosi, M Flemming, F Morady, S A Strickberger.   

Abstract

BACKGROUND: The effect of general anesthesia on defibrillation efficacy in humans is not known. The purpose of this study was to determine the effect of general anesthesia on the defibrillation energy requirements in patients undergoing implantation of a pectoral defibrillator. METHODS AND
RESULTS: Nineteen consecutive patients who underwent defibrillator implantation under general anesthesia were prospectively compared to 16 consecutive patients who underwent defibrillator implantation by the same physicians, using similar devices, at another hospital under conscious sedation. Pre-discharge testing was performed 1.4 +/- 1.0 days after implant using sedation in both groups. The defibrillation energy requirement was determined using the same predefined step-down protocol (15, 10, 8, 5, 3, 1 J) at the time of implantation and during pre-discharge testing. The clinical characteristics of the patients were similar between groups. There was no significant difference in the mean implant defibrillation energy requirement compared to the mean pre-discharge defibrillation energy requirement in either the general anesthesia group (8.5 +/- 4.7 vs. 8.4 +/- 3.4 J; p = 0.9) or in the conscious sedation group (9.4 +/- 3.9 vs. 9.0 +/- 3.8 J; p = 0.7).
CONCLUSIONS: When compared to conscious sedation, general anesthesia with mechanical ventilation has no significant effect on defibrillation efficacy in patients undergoing defibrillator implantation.

Entities:  

Mesh:

Year:  1999        PMID: 10525247     DOI: 10.1023/a:1009831802964

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  26 in total

1.  Temporal stability of defibrillation thresholds with an active pectoral lead system.

Authors:  M R Olsovsky; M A Pelini; S R Shorofsky; M R Gold
Journal:  J Cardiovasc Electrophysiol       Date:  1998-03

2.  Utility of predischarge and one-month transvenous implantable defibrillator tests.

Authors:  J J Goldberger; G Horvath; S Inbar; A H Kadish
Journal:  Am J Cardiol       Date:  1997-03-15       Impact factor: 2.778

3.  High-dose lidocaine does not affect defibrillation efficacy: implications for defibrillation mechanisms.

Authors:  M R Ujhelyi; J J Sims; A W Miller
Journal:  Am J Physiol       Date:  1998-04

4.  [Interaction of frequency-adaptive pacemakers and anesthetic management. Discussion of current literature and two case reports].

Authors:  G von Knobelsdorff; M Goerig; H Nägele; J Scholz
Journal:  Anaesthesist       Date:  1996-09       Impact factor: 1.041

5.  Comparison of etomidate, ketamine, midazolam, propofol, and thiopental on function and metabolism of isolated hearts.

Authors:  D F Stowe; Z J Bosnjak; J P Kampine
Journal:  Anesth Analg       Date:  1992-04       Impact factor: 5.108

6.  Effects of lidocaine on defibrillation threshold in the pig: evidence of anesthesia related increase.

Authors:  A Natale; D L Jones; Y H Kim; G J Klein
Journal:  Pacing Clin Electrophysiol       Date:  1991-08       Impact factor: 1.976

7.  The defibrillation threshold: a comparison of anesthetics and measurement methods.

Authors:  R M Gill; R J Sweeney; P R Reid
Journal:  Pacing Clin Electrophysiol       Date:  1993-04       Impact factor: 1.976

8.  Effects of sevoflurane, isoflurane, and halothane on mechanical and electrophysiologic properties of canine myocardium.

Authors:  N Hatakeyama; Y Ito; Y Momose
Journal:  Anesth Analg       Date:  1993-06       Impact factor: 5.108

9.  Effect of pentobarbital anesthesia on ventricular defibrillation threshold in dogs.

Authors:  C F Babbs
Journal:  Am Heart J       Date:  1978-03       Impact factor: 4.749

10.  Lidocaine causes a reversible, concentration-dependent increase in defibrillation energy requirements.

Authors:  P Dorian; E S Fain; J M Davy; R A Winkle
Journal:  J Am Coll Cardiol       Date:  1986-08       Impact factor: 24.094

View more
  2 in total

1.  Anaesthetic drugs and defibrillation threshold testing.

Authors:  Sony Jacob; Aril E Abraham; George McKelvey
Journal:  Drugs       Date:  2008       Impact factor: 9.546

2.  Safe and effective use of conscious sedation for defibrillation threshold testing during ICD implantation.

Authors:  Ahmed Al Fagih; Hitham Al Shurafa; Saleh Al Ghamdi; Khaled Dagriri; Ayman Al Khadra
Journal:  J Saudi Heart Assoc       Date:  2010-08-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.