Abstract

BACKGROUND: A new portable spirometer, the Spirophone, has been developed that records a subject's blow and can then transmit all the data by telephone to a receiving centre for analysis and comment. Tests of this device were undertaken to determine its accuracy and reliability.
METHODS: The performance of the Spirophone was tested using computer generated wave forms, by delivering blows from calibrated syringes at different flows, and by comparing subjects' blows with those recorded with a commercial spirometer.
RESULTS: Using computer generated wave forms all lung function indices were accurate to within 1% and blows delivered from calibrating syringes were accurate to within 5%. When subjects performed repeated forced vital capacity (FVC) manoeuvres there were no significant differences between lung function indices recorded with the Spirophone and with a commercial spirometer. With the Spirophone and commercial spirometer in series the FVC and forced expiratory volume in one second (FEV1) were within 5% of each other in nine out of 10 healthy subjects.
CONCLUSION: The Spirophone recorded maximal forced expiratory manoeuvres with acceptable accuracy, reliability, and reproducibility, and this system offers the ability to monitor a patient's lung function at a centre remote from the patient.

Entities:  

Mesh:

Year:  1996        PMID: 8733494      PMCID: PMC1090677          DOI: 10.1136/thx.51.4.407

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  22 in total

1.  Heated Fleisch pneumotachometer: a calibration procedure.

Authors:  S Z Turney; W Blumenfeld
Journal:  J Appl Physiol       Date:  1973-01       Impact factor: 3.531

2.  Linearity and temperature control of the Fleisch pneumotachograph.

Authors:  M R Miller; A C Pincock
Journal:  J Appl Physiol (1985)       Date:  1986-02

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  Evaluation of a new ambulatory spirometer for measuring forced expiratory volume in one second and peak expiratory flow rate.

Authors:  D P Johns; M Abramson; G Bowes
Journal:  Am Rev Respir Dis       Date:  1993-05

5.  A procedure for using peak expiratory flow rate data to increase the predictability of asthma episodes.

Authors:  P S Taplin; T L Creer
Journal:  J Asthma Res       Date:  1978-10

6.  Assessing severity of asthma with Wright peak-flow meter.

Authors:  A B Murray; D F Hardwick; G E Pirie; B M Fraser
Journal:  Lancet       Date:  1977-03-26       Impact factor: 79.321

7.  Evaluating commercially available spirometers.

Authors:  R M Gardner; J L Hankinson; B J West
Journal:  Am Rev Respir Dis       Date:  1980-01

8.  The effect of temperature on recording spirograms.

Authors:  A C Pincock; M R Miller
Journal:  Am Rev Respir Dis       Date:  1983-11

9.  National trends in the morbidity and mortality of asthma in the US. Prevalence, hospitalization and death from asthma over two decades: 1965-1984.

Authors:  R Evans; D I Mullally; R W Wilson; P J Gergen; H M Rosenberg; J S Grauman; F M Chevarley; M Feinleib
Journal:  Chest       Date:  1987-06       Impact factor: 9.410

10.  Can patients keep their own peak-flow records reliably?

Authors:  M R Hetzel; I P Williams; R M Shakespeare
Journal:  Lancet       Date:  1979-03-17       Impact factor: 79.321

View more
  2 in total

1.  Telespirometry for home monitoring of pulmonary function.

Authors:  G Izbicki; S Abboud
Journal:  J R Soc Med       Date:  1999-03       Impact factor: 5.344

2.  Benefits of Telemonitoring of Pulmonary Function-3-Month Follow-Up of Home Electronic Spirometry in Patients with Duchenne Muscular Dystrophy.

Authors:  Eliza Wasilewska; Agnieszka Sobierajska-Rek; Sylwia Małgorzewicz; Mateusz Soliński; Ewa Jassem
Journal:  J Clin Med       Date:  2022-02-06       Impact factor: 4.241

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.