Literature DB >> 8732813

Blood glucose reagent strip tests in the operating room: influence of hematocrit, partial pressure of oxygen, and blood glucose level--a comparison of the BM-test 1-44, BM-Accutest, and Satellite G reagent strip systems.

M H Cross1, D G Brown.   

Abstract

OBJECTIVE: The objective of our study was to assess the influence of hematocrit (HCT), partial pressure of oxygen (PO2), and blood glucose level upon results obtained with three different blood glucose reagent strip tests used in conjunction with the appropriate meter: BM-Test 1-44, BM-Accutest, and Satellite G.
METHODS: Our study was designed as a consecutive sample study of patients undergoing coronary artery surgery. The setting was the hospital theater and intensive care unit. We conducted blood analysis for HCT, PO2 and blood glucose on 20 consecutive patients undergoing coronary artery surgery using three blood glucose reagent strip testing systems and a laboratory analysis of plasma glucose.
RESULTS: All three blood glucose reagent strip tests showed a significant bias when compared with plasma glucose: BM-Test 1-44, 0.89 mmol/L; BM-Accutest, -1.27 mmol/L; Satellite G, 0.75 mmol/L (p < 0.05). The error found when using the Satellite G system was worse than that of either of the other two systems. Results obtained with the BM-Accutest strips were unaffected by PO2 (p = 0.745). Blood glucose value and HCT both had an influence on the results of all three blood glucose strip systems.
CONCLUSIONS: Caution must be taken when using reagent strip systems in the operating room or intensive care setting because, of the three systems tested, all showed a significant bias, all were influenced by blood glucose level and HCT, and only the BM-Accutest reagent strips used with the Accutrend meter was unaffected by PO2.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8732813     DOI: 10.1007/bf02025307

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  10 in total

1.  Stick testing.

Authors:  V Marks
Journal:  BMJ       Date:  1991-03-02

2.  The effect of haematocrit on reagent strip tests for glucose.

Authors:  K Wiener
Journal:  Diabet Med       Date:  1991 Feb-Mar       Impact factor: 4.359

Review 3.  Special issues: glucose and the brain.

Authors:  F E Sieber; R J Traystman
Journal:  Crit Care Med       Date:  1992-01       Impact factor: 7.598

4.  How well do nurses perform blood glucose analyses at the diabetic clinic? An assessment using the Reflotron analyser.

Authors:  L A Smith; J McNaught; A S Hutchinson
Journal:  Ann Clin Biochem       Date:  1990-03       Impact factor: 2.057

5.  Laboratory assessment of three new monitors of blood glucose: Accu-Chek II, Glucometer II, and Glucoscan 2000.

Authors:  K E Brooks; N Rawal; A R Henderson
Journal:  Clin Chem       Date:  1986-12       Impact factor: 8.327

6.  Plasma or whole blood glucose?

Authors:  B Morrison; A Fleck
Journal:  Clin Chim Acta       Date:  1973-05-18       Impact factor: 3.786

7.  Consensus statement on self-monitoring of blood glucose.

Authors: 
Journal:  Diabetes Care       Date:  1987 Jan-Feb       Impact factor: 19.112

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 9.  Whole blood glucose: what are we actually measuring?

Authors:  K Wiener
Journal:  Ann Clin Biochem       Date:  1995-01       Impact factor: 2.057

10.  New approach to technical and clinical evaluation of devices for self-monitoring of blood glucose.

Authors:  T Koschinsky; K Dannehl; F A Gries
Journal:  Diabetes Care       Date:  1988-09       Impact factor: 19.112

  10 in total
  1 in total

1.  Assessment of analytical performance of glucose meter in pediatric age group at tertiary care referral hospital.

Authors:  Mohit Vijay Rojekar; Vandana Kumawat; Jayesh Panot; Surekha Khedkar; Arati Adhe-Rojekar
Journal:  J Diabetes Metab Disord       Date:  2017-09-13
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.