A new radiotherapy surface dose detector:the MOSFET.

M J Butson1, A Rozenfeld, J N Mathur, M Carolan, T P Wong, P E Metcalfe.   

Abstract

Radiotherapy x-ray and electron beam surface doses are accurately measurable by use of a MOS-FET detector system. The MOSFET (Metal Oxide Semiconductor Field Effect Transistor) is approximately 200-microns in diameter and consists of a 0.5-microns Al electrode on top of a 1-microns SiO2 and 300-microns Si substrate. Results for % surface dose were within +/- 2% compared to the Attix chamber and within +/- 3% of TLD extrapolation results for normally incident beams. Detectors were compared using different energies, field size, and beam modifying devices such as block trays and wedges. Percentage surface dose for 10 x 10-cm and 40 x 40-cm field size for 6-MV x rays at 100-cm SSD using the MOSFET were 16% and 42% of maximum, respectively. Factors such as its small size, immediate retrieval of results, high accuracy attainable from low applied doses, and as the MOSFET records its dose history make it a suitable in vivo dosimeter where surface and skin doses need to be determined. This can be achieved within part of the first fraction of dose (i.e., only 10 cGy is required.)

Entities:  

Mesh:

Year:  1996        PMID: 8724737     DOI: 10.1118/1.597702

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Effect of a carbon fiber tabletop on the surface dose and attenuation for high-energy photon beams.

Authors:  Tülay P Meydanci; Gönül Kemikler
Journal:  Radiat Med       Date:  2008-11-22

2.  In vivo real-time rectal wall dosimetry for prostate radiotherapy.

Authors:  Nicholas Hardcastle; Dean L Cutajar; Peter E Metcalfe; Michael L F Lerch; Vladimir L Perevertaylo; Wolfgang A Tomé; Anatoly B Rosenfeld
Journal:  Phys Med Biol       Date:  2010-07-07       Impact factor: 3.609

3.  A novel phantom model for mouse tumor dose assessment under MV beams.

Authors:  Michael S Gossman; Indra J Das; Subhash C Sharma; Jeffrey P Lopez; Candace M Howard; Pier Claudio
Journal:  Health Phys       Date:  2011-12       Impact factor: 1.316

4.  Is the in vivo dosimetry with the OneDosePlusTM system able to detect intra-fraction motion? A retrospective analysis of in vivo data from breast and prostate patients.

Authors:  Maria Daniela Falco; Marco D'Andrea; Alessia Lo Bosco; Mauro Rebuzzi; Elisabetta Ponti; Barbara Tolu; Grazia Tortorelli; Rosaria Barbarino; Luana Di Murro; Riccardo Santoni
Journal:  Radiat Oncol       Date:  2012-06-20       Impact factor: 3.481

5.  Peripheral doses in patients undergoing Cyberknife treatment for intracranial lesions. A single centre experience.

Authors:  Vassiliki Vlachopoulou; Christos Antypas; Harry Delis; Argyrios Tzouras; Nikolaos Salvaras; Dimitrios Kardamakis; George Panayiotakis
Journal:  Radiat Oncol       Date:  2011-11-14       Impact factor: 3.481

6.  A novel approach towards development of real time chemical dosimetry using pulsating sensor-based instrumentation.

Authors:  N Malathi; P Sahoo; K Praveen; N Murali
Journal:  J Radioanal Nucl Chem       Date:  2013-05-14       Impact factor: 1.371

7.  Characterization of MOSFET dosimeter angular dependence in three rotational axes measured free-in-air and in soft-tissue equivalent material.

Authors:  Juha Koivisto; Timo Kiljunen; Jan Wolff; Mika Kortesniemi
Journal:  J Radiat Res       Date:  2013-03-21       Impact factor: 2.724

8.  Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams.

Authors:  A Sathish Kumar; S D Sharma; B Paul Ravindran
Journal:  J Med Phys       Date:  2014-07

9.  Characterization of MOSFET dosimeters for low-dose measurements in maxillofacial anthropomorphic phantoms.

Authors:  Juha H Koivisto; Jan E Wolff; Timo Kiljunen; Dirk Schulze; Mika Kortesniemi
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.