Pull-out strength of pedicle hooks with fixation screws: influence of screw length and angulation.

U Berlemann1, P A Cripton, L Rincon, L P Nolte, F Schläpfer.   

Abstract

The pull-out force of thoracic spinal pedicle hooks secured by long fixation screws engaging the posterior portion of the vertebral endplate was measured. The performance of these hooks was compared with that of hooks using a shorter screw and different screw orientation such that the vertebral endplates were not perforated. The longer and differently angulated screws, engaging the endplate, significantly enhanced the fixation potential of the hooks.

Mesh:

Year:  1996        PMID: 8689420     DOI: 10.1007/bf00307831

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  3 in total

1.  The holding power of orthopedic screws in vivo.

Authors:  J Schatzker; R Sanderson; J P Murnaghan
Journal:  Clin Orthop Relat Res       Date:  1975-05       Impact factor: 4.176

Review 2.  Anatomic and technical considerations of pedicle screw fixation.

Authors:  J N Weinstein; B L Rydevik; W Rauschning
Journal:  Clin Orthop Relat Res       Date:  1992-11       Impact factor: 4.176

3.  New means in spinal pedicle hook fixation. A biomechanical evaluation.

Authors:  U Berlemann; P Cripton; L P Nolte; K Lippuner; F Schläpfer
Journal:  Eur Spine J       Date:  1995       Impact factor: 3.134

  3 in total
  6 in total

1.  A biomechanical analysis of the self-retaining pedicle hook device in posterior spinal fixation.

Authors:  Wilbert van Laar; Rinse J Meester; Theo H Smit; Barend J van Royen
Journal:  Eur Spine J       Date:  2007-01-03       Impact factor: 3.134

2.  Ideal entry point for the thoracic pedicle screw during the free hand technique.

Authors:  Kook Jin Chung; Seung Woo Suh; Sameer Desai; Hae Ryong Song
Journal:  Int Orthop       Date:  2007-04-17       Impact factor: 3.075

3.  Which salvage fixation technique is best for the failed initial screw fixation at the cervicothoracic junction? A biomechanical comparison study.

Authors:  Jae Taek Hong; Takigawa Tomoyuki; Ashish Jain; Alejandro A Espinoza Orías; Nozomu Inoue; Howard S An
Journal:  Eur Spine J       Date:  2017-07-27       Impact factor: 3.134

Review 4.  Pedicle screw fixation in spinal disorders: a European view.

Authors:  N Boos; J K Webb
Journal:  Eur Spine J       Date:  1997       Impact factor: 3.134

Review 5.  Hydroxyapatite Use in Spine Surgery-Molecular and Clinical Aspect.

Authors:  Jakub Litak; Wojciech Czyzewski; Michał Szymoniuk; Bartlomiej Pastuszak; Joanna Litak; Grzegorz Litak; Cezary Grochowski; Mansur Rahnama-Hezavah; Piotr Kamieniak
Journal:  Materials (Basel)       Date:  2022-04-15       Impact factor: 3.748

6.  Clinical significance of superior articular process as a reference point for free-hand pedicle screw insertion in thoracic spine.

Authors:  Tae Hoon Kim; Suk Ha Lee; Jae Hyuk Yang; Jae Young Hong; Seung Woo Suh
Journal:  Medicine (Baltimore)       Date:  2018-02       Impact factor: 1.817

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.