Analysis of load transfer and stress distribution by an implant-supported fixed partial denture.

D Assif1, B Marshak, A Horowitz.   

Abstract

This study simultaneously examined the load transfer and stress distribution by an implant-supported fixed partial denture. A mandibular implant framework with implants connected to the abutments was embedded in a three-dimensional photoelastic model of a mandible. Strain gauges were attached on the superior surface of the framework, and a vertical load of 7.5 kg was applied to seven points on the framework. The measurements derived from this simulation revealed that (1) there was a direct proportion between the stress distribution in the metal framework and stresses created in the supporting structure around the implants; (2) the mode of load transfer and stress distribution was directly proportional to the distance of the components from the loading point; and (3) when the cantilever was loaded, the major part of the stress was distributed within the cantilever in the connection to the distal abutment. In this simulation, stress was distributed over the two, or maximum three, closest implants with the distal implant the most stressed.

Entities:  

Mesh:

Year:  1996        PMID: 8648576     DOI: 10.1016/s0022-3913(96)90486-2

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  4 in total

1.  Standard of disocclusion in complete dentures supported by implants without free distal ends: analysis by the finite elements method.

Authors:  Gustavo Diniz Greco; Estevam Barbosa de Las Casas; Tulimar P Machado Cornacchia; Cláudia Silami de Magalhães; Allyson Nogueira Moreira
Journal:  J Appl Oral Sci       Date:  2012-02       Impact factor: 2.698

2.  Biomechanical analysis of the stresses generated by different disocclusion patterns in an implant-supported mandibular complete denture.

Authors:  Gustavo Diniz Greco; Wellington Corrêa Jansen; Janis Landre Junior; Paulo Isaías Seraidarian
Journal:  J Appl Oral Sci       Date:  2009 Sep-Oct       Impact factor: 2.698

3.  Stress on external hexagon and Morse taper implants submitted to immediate loading.

Authors:  Caroline H Odo; Marcele J Pimentel; Rafael L X Consani; Marcelo F Mesquita; Mauro A A Nóbilo
Journal:  J Oral Biol Craniofac Res       Date:  2015-07-29

4.  Effect of cantilever length and alloy framework on the stress distribution in peri-implant area of cantilevered implant-supported fixed partial dentures.

Authors:  Valdey Suedam; Rafael Tobias Moretti Neto; Edson Antonio Capello Sousa; José Henrique Rubo
Journal:  J Appl Oral Sci       Date:  2016-04       Impact factor: 2.698

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.