M J Gains1, K M Grodecki, R M Jacobs, D Dyson, R A Foster.
Abstract
The precision and accuracy of an indirect oscillometric blood pressure measurement technique (Dinamap 8100) was assessed in 11 anesthetized Beagle dogs weighing 8 to 11.5 kg. Direct blood pressure measurements were made by catheterization of the lingual artery, and simultaneous indirect measurements were determined by placing a cuff over the median artery (midradial area). Blood pressure measurements at 2 different planes of anesthesia (light and deep) were recorded in triplicate. At a light plane of anesthesia, the Dinamap 8100 underestimated diastolic and mean arterial pressure, and at a deep anesthetic plane overestimated systolic pressure. The indirect technique had good repeatability of systolic pressures. Regression analysis for the 2 techniques showed excellent correlation (r = 0.93). The results indicate that the indirect oscillometric blood pressure measurement technique provides a good estimate of systolic, diastolic, and mean arterial pressure in dogs weighing 8-11.5 kg.Entities:
Mesh:
Year: 1995 PMID: 8521360 PMCID: PMC1263773
Source DB: PubMed Journal: Can J Vet Res ISSN: 0830-9000 Impact factor: 1.310