| Literature DB >> 843296 |
G Bergmann, R Kölbel, A Rohlmann.
Abstract
In joint replacement the fixation of implants to bone is loaded dynamically. The fatique strength of the bonding of PMMA to cancellous bone was determined for cyclic tensile loading. It was found to be 35% of the static tensile strength and greater than 0.6 Nmm-2 within a 95%-safety-range.Entities:
Mesh:
Substances:
Year: 1977 PMID: 843296 DOI: 10.1007/BF00415210
Source DB: PubMed Journal: Arch Orthop Unfallchir ISSN: 0003-9330