Literature DB >> 8346488

Improvement in sodium cromoglycate delivery from a spacer device by use of an antistatic lining, immediate inhalation, and avoiding multiple actuations of drug.

C O'Callaghan1, J Lynch, M Cant, C Robertson.   

Abstract

BACKGROUND: Aerosols generated from metered dose inhalers may be highly charged. The aim of this study was to determine whether lining the walls of a polycarbonate spacer device with an antistatic agent would result in an increase in drug output. The effects of multiple actuations of drug into the spacer device and increasing residence time of drug within the spacer were also determined.
METHODS: The amount of sodium cromoglycate contained in particles of various size available for inhalation (per 5 mg actuation) from a 750 ml polycarbonate spacer was determined by impinger measurement and spectrophotometric assay.
RESULTS: Lining the spacer with an antistatic agent increased the mean (SD) amount of sodium cromoglycate in particles < 5 microns available for inhalation (per 5 mg actuation) by 244% from (0.59 (0.03) to 1.44 (0.2) mg). When there was a 20 second interval between actuation into the spacer device and inhalation, sodium cromoglycate available for inhalation in particles < 5 micrograms decreased by 67% (from 0.59 (0.03) mg to 0.2 (0.01) mg). Use of the spacer device increased sodium cromoglycate available for inhalation in respirable particles (< 5 microns) by 18% compared with direct delivery by metered dose inhaler. Multiple actuations into the spacer decreased the amount of sodium cromoglycate available for inhalation in particles < 5 microns by 31% after two actuations and 56% after three acutations.
CONCLUSIONS: Multiple actuations of sodium cromoglycate into a spacer device before inhalation should be avoided, and inhalation from spacer devices should take place immediately after actuation to ensure maximum dose. Lining of a standard spacer device with an antistatic agent significantly increased output of sodium cromoglycate. This may have implications for improved therapeutic response and drug cost.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8346488      PMCID: PMC464576          DOI: 10.1136/thx.48.6.603

Source DB:  PubMed          Journal:  Thorax        ISSN: 0040-6376            Impact factor:   9.139


  4 in total

1.  How much aerosol reaches the lungs of wheezy infants and toddlers?

Authors:  B Salmon; N M Wilson; M Silverman
Journal:  Arch Dis Child       Date:  1990-04       Impact factor: 3.791

2.  Spacer device with face mask attachment for giving bronchodilators to infants with asthma.

Authors:  C O'Callaghan; A D Milner; A Swarbrick
Journal:  BMJ       Date:  1989-01-21

3.  On the deposition of unipolarly charged particles in the human respiratory tract.

Authors:  C Melandri; V Prodi; G Tarroni; M Formignani; T De Zaiacomo; G F Bompane; G Maestri
Journal:  Inhaled Part       Date:  1975-09

4.  Deposition of pressurized suspension aerosols inhaled through extension devices.

Authors:  S P Newman; F Morén; D Pavia; F Little; S W Clarke
Journal:  Am Rev Respir Dis       Date:  1981-09
  4 in total
  27 in total

1.  Randomised controlled study of clinical efficacy of spacer therapy in asthma with regard to electrostatic charge.

Authors:  E Dompeling; A M Oudesluys-Murphy; H M Janssens; W Hop; J G Brinkman; R N Sukhai; J C de Jongste
Journal:  Arch Dis Child       Date:  2001-02       Impact factor: 3.791

2.  Reducing electrostatic charge on spacer devices and bronchodilator response.

Authors:  J H Wildhaber; G W Waterer; G L Hall; Q A Summers
Journal:  Br J Clin Pharmacol       Date:  2000-09       Impact factor: 4.335

3.  Asthma drug delivery devices for children.

Authors:  C O'Callaghan; P W Barry
Journal:  BMJ       Date:  2000-03-11

4.  Aerosol profile extracted from spacers as a determinant of actual dose.

Authors:  Sylvia Verbanck; Chris Vervaet; Daniël Schuermans; Walter Vincken
Journal:  Pharm Res       Date:  2004-12       Impact factor: 4.200

5.  Effect of oropharyngeal length in drug lung delivery via suspension pressurized metered dose inhalers.

Authors:  T Ehtezazi; D R Allanson; I D Jenkinson; C O'Callaghan
Journal:  Pharm Res       Date:  2006-05-26       Impact factor: 4.200

Review 6.  Deposition and effects of inhaled corticosteroids.

Authors:  Stephen P Newman
Journal:  Clin Pharmacokinet       Date:  2003       Impact factor: 6.447

Review 7.  Optimizing inhaled drug delivery in patients with asthma.

Authors:  C Jackson; B Lipworth
Journal:  Br J Gen Pract       Date:  1995-12       Impact factor: 5.386

8.  Particle size selection device for use with the Turbohaler.

Authors:  M L Everard; S G Devadason; P N Le Souef
Journal:  Thorax       Date:  1996-05       Impact factor: 9.139

9.  Effect of multiple actuations, delayed inhalation and antistatic treatment on the lung bioavailability of salbutamol via a spacer device.

Authors:  D J Clark; B J Lipworth
Journal:  Thorax       Date:  1996-10       Impact factor: 9.139

Review 10.  Spacer devices for metered dose inhalers.

Authors:  Stephen P Newman
Journal:  Clin Pharmacokinet       Date:  2004       Impact factor: 6.447

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.