Literature DB >> 8331121

Tibial high-density polyethylene wear in conforming tibiofemoral prostheses.

P Plante-Bordeneuve1, M A Freeman.   

Abstract

We have studied 27 tibial prostheses retrieved from knee replacements after 1 to 9 years. In 22 the femoral components were of cobalt-chrome, in five polyacetal. The design of the components gave a nominal contact area of 320 mm2 on each condyle. The tibial component was of high-density polyethylene (HDP) at least 6 mm thick, and not heat-treated. In the metal/HDP prostheses the average wear rate was 0.025 mm/year. The relative wear on the medial and lateral sides was related to the leg axis. None of the retrieved prostheses showed any severe disruption of their surface. The polyacetal/HDP prostheses showed similar wear with a statistically insignificant trend towards slower penetration. We conclude that the rate of wear of HDP in a conforming tibiofemoral bearing with a fixed tibial component at least 6 mm thick and not heat-treated is slow enough to be safe in clinical practice.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8331121

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  7 in total

1.  Toward the interpretation of the combined effect of size and body weight on the tribological performance of total knee prostheses.

Authors:  Santina Battaglia; Paola Taddei; Silvia Tozzi; Alessandra Sudanese; Saverio Affatato
Journal:  Int Orthop       Date:  2014-02-26       Impact factor: 3.075

2.  Is isolated insert exchange a valuable choice for polyethylene wear in metal-backed unicompartmental knee arthroplasty?

Authors:  Alexandre Lunebourg; Sébastien Parratte; Alexandre Galland; François Lecuire; Matthieu Ollivier; Jean-Noël Argenson
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-10-25       Impact factor: 4.342

3.  An autonomous mathematical reconstruction to effectively measure volume loss on retrieved polyethylene tibial inserts.

Authors:  Christopher B Knowlton; Markus A Wimmer
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2012-08-22       Impact factor: 3.368

4.  Influence of tibiofemoral congruency design on the wear of patient-specific unicompartmental knee arthroplasty using finite element analysis.

Authors:  Y-G Koh; K-M Park; H-Y Lee; K-T Kang
Journal:  Bone Joint Res       Date:  2019-04-02       Impact factor: 5.853

5.  Unique relationship between osteophyte and femoral-tibia component size mismatch in determining polyethylene wear in primary total knee arthroplasty: a case report.

Authors:  Manjunath Ramappa; Andrew Port
Journal:  J Med Case Rep       Date:  2009-02-10

6.  Short-Term Results of Total Knee Arthroplasty with Anterior-Posterior Glide LCS Mobile-Bearing System.

Authors:  Kyung Taek Kim; Min Soo Kang; Young Hoon Lim; Jin Woo Park; Lih Wang
Journal:  Knee Surg Relat Res       Date:  2014-08-29

7.  Immediate Postoperative Portable Radiograph After Total Knee Replacements: A Necessity or a Burden?

Authors:  Abdulla Aljawder; Dana Alomran; Mohammed Alayyoub; Fahad Alkhalifa
Journal:  Open Orthop J       Date:  2018-05-31
  7 in total