Relation between upper limit of vulnerability and defibrillation threshold in humans.

P S Chen1, G K Feld, J M Kriett, M M Mower, R Y Tarazi, R P Fleck, C D Swerdlow, E S Gang, R M Kass.   

Abstract

BACKGROUND: In the canine model, an upper limit of shock strength exists that can induce ventricular fibrillation during the vulnerable period of the cardiac cycle. This shock strength (the upper limit of vulnerability) closely correlates with the defibrillation threshold and supports the "upper limit of vulnerability" hypothesis of defibrillation. It is not known whether an upper limit of vulnerability exists in humans or whether this limit correlates with the defibrillation threshold. METHODS AND
RESULTS: In 13 patients undergoing implantable cardioverter-defibrillator implantation, the shock strengths associated with a 50% probability of reaching the upper limit of vulnerability (ULV50) and a 50% probability of reaching the defibrillation threshold (DFT50) were determined by the up-down algorithm. The ULV50 was determined only for the mid-upslope of the positive T waves and for the mid-downslope of the negative T waves. No major complications occurred during surgery. An upper limit of vulnerability was demonstrated in each patient. The ULV50 was 300 +/- 138 V or 6.8 +/- 5.8 J, which was significantly lower than the DFT50 of 347 +/- 167 V (p = 0.038) or 9.1 +/- 7.3 J (p = 0.013). The correlation between the ULV50 and the DFT50 was significant (r = 0.90, p < 0.001 for voltage; r = 0.93, p < 0.001 for energy).
CONCLUSIONS: An upper limit of vulnerability is present in humans. There is a significant correlation between the ULV50 and the DFT50, and the ULV50 is significantly lower than the DFT50.

Entities:  

Mesh:

Year:  1993        PMID: 8319332     DOI: 10.1161/01.cir.88.1.186

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  11 in total

Review 1.  Mechanisms of defibrillation.

Authors:  Derek J Dosdall; Vladimir G Fast; Raymond E Ideker
Journal:  Annu Rev Biomed Eng       Date:  2010-08-15       Impact factor: 9.590

2.  Differences between left and right ventricular chamber geometry affect cardiac vulnerability to electric shocks.

Authors:  Blanca Rodríguez; Li Li; James C Eason; Igor R Efimov; Natalia A Trayanova
Journal:  Circ Res       Date:  2005-06-23       Impact factor: 17.367

3.  The role of mechanoelectric feedback in vulnerability to electric shock.

Authors:  Weihui Li; Viatcheslav Gurev; Andrew D McCulloch; Natalia A Trayanova
Journal:  Prog Biophys Mol Biol       Date:  2008-02-16       Impact factor: 3.667

4.  Correlation of acute and chronic defibrillation threshold with upper limit of vulnerability determined in normal sinus rhythm.

Authors:  U Birgersdotter-Green; K Undesser; O Fujimura; G K Feld; R M Kass; W J Mandel; C T Peter; P S Chen
Journal:  J Interv Card Electrophysiol       Date:  1999-07       Impact factor: 1.900

Review 5.  New insights into defibrillation of the heart from realistic simulation studies.

Authors:  Natalia A Trayanova; Lukas J Rantner
Journal:  Europace       Date:  2014-05       Impact factor: 5.214

6.  Induction of ventricular fibrillation by T wave shocks: observations from monophasic action potential recordings.

Authors:  R K Shepard; M A Wood; D Dan; H F Clemo; D M Gilligan; K A Ellenbogen
Journal:  J Interv Card Electrophysiol       Date:  1999-12       Impact factor: 1.900

7.  Chronaxie of defibrillation: a pathway toward further optimization of defibrillation waveform?

Authors:  Igor R Efimov
Journal:  J Cardiovasc Electrophysiol       Date:  2008-10-14

8.  Successful implantation of cardiac defibrillators without induction of ventricular fibrillation using upper limit of vulnerability testing.

Authors:  Ulrika Birgersdotter Green; Ashok Garg; Fawzia Al-Kandari; GilAnthony Ungab; Linda Tone; Gregory K Feld
Journal:  J Interv Card Electrophysiol       Date:  2003-02       Impact factor: 1.900

9.  Upper limit of vulnerability and heterogeneity.

Authors:  Peng-Sheng Chen; Shien-Fong Lin
Journal:  Heart Rhythm       Date:  2008-12-03       Impact factor: 6.343

10.  High defibrillation threshold: the science, signs and solutions.

Authors:  Sony Jacob; Victorio Pidlaoan; Jaspreet Singh; Aditya Bharadwaj; Mehul B Patel; Antonio Carrillo
Journal:  Indian Pacing Electrophysiol J       Date:  2010-01-07