Correlation of acute and chronic defibrillation threshold with upper limit of vulnerability determined in normal sinus rhythm.

U Birgersdotter-Green1, K Undesser, O Fujimura, G K Feld, R M Kass, W J Mandel, C T Peter, P S Chen.   

Abstract

BACKGROUND: The upper limit of vulnerability (ULV) is the stimulus strength above which ventricular fibrillation cannot be induced, even when the stimulus occurs during the vulnerable period of the cardiac cycle. Determination of ULV using T-wave shocks during ventricular pacing has been shown to closely correlate with the defibrillation threshold (DFT) at ICD implantation. However, there are no data correlating ULV determined in sinus rhythm at ICD implantation, with DFT determined at implantation or during long-term follow-up. This is of clinical importance since ULV may be used to estimate DFT during ICD implantation, both during ventricular pacing or sinus rhythm. METHODS AND
RESULTS: Twenty-one patients receiving a transvenous ICD system were studied prospectively. There were 16 males and 5 females, mean age 68 +/- 15 years, with mean ejection fraction 37.4 +/- 17.4%. All had structural heart disease. The ULV was defined as the lowest energy that did not induce ventricular fibrillation with shocks at 0, 20 and 40ms before the peak of the T-wave, using a step-down protocol. The initial energy tested was 15J and the lowest energy 2J. DFT was determined following a similar step-down protocol. The DFT was defined as the lowest energy that successfully defibrillated the ventricles. The linear correlation coefficient between ULV and DFT was r = 0.73 (p < 0.001). At implant, mean ULV was 9.2 +/- 5J, not statistically different from mean DFT 9.4 +/- 4J. ULV plus 5J successfully defibrillated 19 of 21 patients. During long-term follow-up of 10.1 +/- 1.8 months in eight patients, DFT was 8.8 +/- 5.8J, not significantly different than the DFT of 7.5 +/- 4.1J or ULV of 8.0 +/- 5.3 at implant.
CONCLUSION: 1) When determined during normal sinus rhythm the ULV significantly correlates with DFT. 2) ULV testing might be used in lieu of standard DFT testing to confirm adequate lead placement thus minimizing or eliminating VF inductions, particularly in hemodynamically unstable patients. 3) Since ULV + 5J has a high probability of successful defibrillation in most patients, programming ICD first shock energy for VF at ULV + 5J may result in lower first shock energies compared to the standard methods of programming first shock energy at twice DFT. CONDENSED ABSTRACT: The purpose of this study was to determine if the upper limit of vulnerability (ULV) determined during normal sinus rhythm correlates with the defibrillation threshold (DFT), as has been previously shown when determined during ventricular pacing. The linear correlation coefficient between the ULV and DFT was r = 0.73 (p < 0.001). Mean ULV at implant was 9.2 +/- 5J, not statistically different from mean DFT of 0.4 +/- 4J. During long-term follow-up of 10.1 +/- 1.8 months in 8 patients, DFT was 8.75 +/- 8J, not significantly different than the DFT of 7.5 +/- 4.1J or ULV of 8.0 +/- 5.3 at implant. Shocks energies of ULV + 5J successfully defibrillated 19 of 21 patients at implant and 8 of 8 at follow-up. This study indicates that the ULV determined in normal sinus rhythm closely correlates with the DFT, and that ULV + 5J defibrillated most patients. ULV testing could be used to predict DFT and reduce or eliminate the need for DFT testing and VF induction. Programming ICD first shock energy for VF to ULV + 5J will result in lower energy than that used with standard DFT testing.

Entities:  

Mesh:

Year:  1999        PMID: 10387143     DOI: 10.1023/a:1009825731592

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  13 in total

1.  Results and realistic expectations with transvenous lead systems.

Authors:  M Block; D Hammel; F Isbruch; M Borggrefe; D Wietholt; T Hachenberg; H H Scheld; G Breithardt
Journal:  Pacing Clin Electrophysiol       Date:  1992-04       Impact factor: 1.976

2.  The study of ventricular fibrillation by threshold determinations.

Authors:  N E SHUMWAY; J A JOHNSON; R J STISH
Journal:  J Thorac Surg       Date:  1957-11

3.  Effects of pacing rate and timing of defibrillation shock on the relation between the defibrillation threshold and the upper limit of vulnerability in open chest dogs.

Authors:  P S Chen; G K Feld; M M Mower; B B Peters
Journal:  J Am Coll Cardiol       Date:  1991-11-15       Impact factor: 24.094

4.  An energy-time analysis of ventricular fibrillation and defibrillation thresholds with internal electrodes.

Authors:  C Lesigne; B Levy; R Saumont; P Birkui; A Bardou; B Rubin
Journal:  Med Biol Eng       Date:  1976-11

5.  [The threshold of synchronous response of the myocardial fibers. Application to the experimental comparison of the efficacy of different forms of electroshock defibrillation].

Authors:  A Fabiato; P Coumel; R Gourgon; R Saumont
Journal:  Arch Mal Coeur Vaiss       Date:  1967-04

6.  Temporal dispersion of recovery of excitability in atrium and ventricle as a function of heart rate.

Authors:  J Han; D Millet; B Chizzonitti; G K Moe
Journal:  Am Heart J       Date:  1966-04       Impact factor: 4.749

7.  Fibrillation threshold of premature ventricular responses.

Authors:  J Han; P D Garcia de Jalon; G K Moe
Journal:  Circ Res       Date:  1966-01       Impact factor: 17.367

8.  Clinical evaluation of the safety of repetitive intraoperative defibrillation threshold testing.

Authors:  R Frame; R Brodman; S Furman; S G Kim; J Roth; K Ferrick; I Hollinger; J Gross; J D Fisher
Journal:  Pacing Clin Electrophysiol       Date:  1992-06       Impact factor: 1.976

9.  Upper limit of vulnerability reliably predicts the defibrillation threshold in humans.

Authors:  C Hwang; C D Swerdlow; R M Kass; E S Gang; W J Mandel; C T Peter; P S Chen
Journal:  Circulation       Date:  1994-11       Impact factor: 29.690

10.  Comparison of the defibrillation threshold and the upper limit of ventricular vulnerability.

Authors:  P S Chen; N Shibata; E G Dixon; R O Martin; R E Ideker
Journal:  Circulation       Date:  1986-05       Impact factor: 29.690

View more
  2 in total

1.  [Single- and dual-chamber ICDs: Are there still significant differences compared to pacemakers with regard to implantation and follow-up?].

Authors:  M Stockburger
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2008-12

2.  Successful implantation of cardiac defibrillators without induction of ventricular fibrillation using upper limit of vulnerability testing.

Authors:  Ulrika Birgersdotter Green; Ashok Garg; Fawzia Al-Kandari; GilAnthony Ungab; Linda Tone; Gregory K Feld
Journal:  J Interv Card Electrophysiol       Date:  2003-02       Impact factor: 1.900

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.