Literature DB >> 8315816

Long-term follow up of ventricular endocardial pacing leads. Complications, electrical performance, and longevity of 561 right ventricular leads.

S Kazama1, K Nishiyama, M Machii, K Tanaka, T Amano, T Nomura, M Ohuchi, S Kasahara, M Nie, A Ishihara.   

Abstract

Five hundred and sixty-one endocardial pacing leads implanted in the right ventricle of 502 patients between 1971 and 1990 were followed for up to 17 years regarding their complications, stimulation threshold behavior, and overall longevity. Lead tip dislodgment occurred in 16 leads (2.9%), in 14 of which dislodgement occurred within 2 months after implantation. The incidence of dislodgement was significantly smaller in tined leads than in nontined leads. Lead conductor fracture occurred in 19 leads (3.4%), in 15 of which fracture occurred within 5 years after implantation. Two particular sites of lead fracture were identified; i.e., one within the pacemaker pocket and the other at a particular point in the subclavian vein between the clavicle and the first rib. Fixation ligature in the former site and the venopuncture point for lead insertion in the latter site are thought to be related to lead fracture. Other complications included insulator break in 3 leads (0.5%), exit block in 7 leads (1.2%), and poor sensing in 2 leads (0.4%). The cumulative survival of leads was 94.1 +/- 2.7% (mean +/- 95% confidence interval) at 5 years, 86.3 +/- 6.3% at 10 years, and 74.2 +/- 14.0% at 15 years after implantation. The minimal stimulation threshold, lead impedance and R wave amplitude were all found to be stable for up to 10 years, and there were no tendencies toward higher stimulation threshold during the observation period of this study.

Entities:  

Mesh:

Year:  1993        PMID: 8315816     DOI: 10.1536/ihj.34.193

Source DB:  PubMed          Journal:  Jpn Heart J        ISSN: 0021-4868


  8 in total

1.  Pacemaker lead entrapment syndrome.

Authors:  Kamal Gupta; Rollo P Villareal; Abdi Rasekh; Ali Massumi
Journal:  Tex Heart Inst J       Date:  2003

2.  Axillary vein puncture using fluoroscopic landmarks: a safe and effective approach for implantable cardioverter defibrillator leads.

Authors:  Federico Migliore; Mariachiara Siciliano; Manuel De Lazzari; Sonia Ferretto; Chiara Dalla Valle; Alessandro Zorzi; Domenico Corrado; Sabino Iliceto; Emanuele Bertaglia
Journal:  J Interv Card Electrophysiol       Date:  2015-05-09       Impact factor: 1.900

3.  Mechanical fatigue resistance of an implantable branched lead system for a distributed set of longitudinal intrafascicular electrodes.

Authors:  A E Pena; S S Kuntaegowdanahalli; J J Abbas; J Patrick; K W Horch; R Jung
Journal:  J Neural Eng       Date:  2017-12       Impact factor: 5.379

4.  Efficacy of ultrasound-guided axillary/subclavian venous approaches for pacemaker and defibrillator lead implantation: a randomized study.

Authors:  Mattia Liccardo; Pasquale Nocerino; Salzano Gaia; Carmine Ciardiello
Journal:  J Interv Card Electrophysiol       Date:  2018-01-15       Impact factor: 1.900

5.  Prospective study to develop surface landmarks for blind axillary vein puncture for permanent pacemaker and defibrillator lead implantation and compare it to available contrast venography guided technique.

Authors:  S Mehrotra; Manoj Kumar Rohit
Journal:  Indian Heart J       Date:  2015-05-13

6.  Possible complications of subclavian crush syndrome.

Authors:  S A M Said; C H J M Ticheler; C M Stassen; A Derks; H T Droste
Journal:  Neth Heart J       Date:  2005-03       Impact factor: 2.380

7.  Complication rates associated with transvenous pacemaker implantation in dogs with high-grade atrioventricular block performed during versus after normal business hours.

Authors:  J L Ward; T C DeFrancesco; S P Tou; C E Atkins; E H Griffith; B W Keene
Journal:  J Vet Intern Med       Date:  2015-01       Impact factor: 3.333

8.  Medial subclavicular musculotendinous complex and insulation break: Rare cause of late pacemaker lead malfunction.

Authors:  Pranab Jyoti Bhattacharyya; Shweta Agrawal; Jogesh Chandra Barkataky; Anjan Kumar Bhattacharyya
Journal:  Indian Heart J       Date:  2015-09-11
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.