Precision and accuracy of stereotactic convergent beam irradiations from a linear accelerator.

G H Hartmann1, B Bauer-Kirpes, C F Serago, W J Lorenz.   

Abstract

PURPOSE: The accuracy and the precision for radiosurgery procedures at linear accelerator facilities were investigated. METHODS AND MATERIALS: The technique of convergent beam irradiation, that is a series of successive isocentric arc irradiations, is specifically considered in this paper. Accuracy and precision depend on a sequence of methods and equipment among which localization of the target, patient alignment, and the dose delivery are the most critical steps. The purpose of the investigation was to quantitatively assess their contribution to the overall accuracy. The definitions and methods used to quantify and control accuracy are described. Measurements were carried out at a phantom to analyze the localization and positioning errors. Errors which may occur with the dose delivery technique were studied by a computer simulation.
RESULTS: The calculations showed that these errors are not the main contributors to the overall accuracy as long as the linac inaccuracies are in the order or less than 1 mm. The accuracy found in the localization and positioning methods was less than 1 mm.
CONCLUSION: It was concluded that an overall accuracy in the order of 1 mm can be obtained also under routine conditions. The great importance of adequate quality control is emphasized.

Entities:  

Mesh:

Year:  1994        PMID: 8276665     DOI: 10.1016/0360-3016(94)90075-2

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  5 in total

1.  Clinical results of a pilot study on stereovision-guided stereotactic radiotherapy and intensity modulated radiotherapy.

Authors:  Shidong Li; Lawrence R Kleinberg; Daniele Rigamonti; Moody D Wharam; Abdul Rashid; Juan Jackson; David Djajaputra; Shenjen He; Tunisia Creasey; Theodore L DeWeese
Journal:  Technol Cancer Res Treat       Date:  2010-12

2.  On the accuracy of isocenter verification with kV imaging in stereotactic radiosurgery.

Authors:  Rolf Wiehle; Hans-Jürgen Koth; Norbert Nanko; Anca-Ligia Grosu; Norbert Hodapp
Journal:  Strahlenther Onkol       Date:  2009-05-15       Impact factor: 3.621

3.  The application of micro-vacuo-certo-contacting ophthalmophanto in X-ray radiosurgery for tumors in an eyeball.

Authors:  Shuying Li; Yunyan Wang; Likuan Hu; Yingchun Liang; Jing Cai
Journal:  Tumour Biol       Date:  2014-08-15

4.  A closer look at the conventional Winston-Lutz test: Analysis in terms of dose.

Authors:  Juan-Francisco Calvo-Ortega; Sandra Moragues-Femenía; Coral Laosa-Bello; Sol San José-Maderuelo; Joan Casals-Farran
Journal:  Rep Pract Oncol Radiother       Date:  2019-07-18

Review 5.  Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.