Temporal lobes: surface MR coil phased-array imaging.

C E Hayes1, J S Tsuruda, C M Mathis.   

Abstract

The authors describe a four-coil surface phased-array coil for magnetic resonance (MR) imaging of the temporal lobes. On the basis of measurements in a phantom, the signal-to-noise ratio of the array is 1.67 times or more better than that obtained with a standard quadrature head coil in the region of the hippocampi. The higher sensitivity of this phased array permits use of imaging pulse sequences for which the head coil is inadequate.

Mesh:

Year:  1993        PMID: 8234726     DOI: 10.1148/radiology.189.3.8234726

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  5 in total

1.  High-sensitivity coil array for head and neck imaging: technical note.

Authors:  R G Henry; N J Fischbein; W P Dillon; D B Vigneron; S J Nelson
Journal:  AJNR Am J Neuroradiol       Date:  2001 Nov-Dec       Impact factor: 3.825

2.  Evaluation of image quality of a 32-channel versus a 12-channel head coil at 1.5T for MR imaging of the brain.

Authors:  P T Parikh; G S Sandhu; K A Blackham; M D Coffey; D Hsu; K Liu; J Jesberger; M Griswold; J L Sunshine
Journal:  AJNR Am J Neuroradiol       Date:  2010-12-16       Impact factor: 3.825

3.  Design and numerical evaluation of a volume coil array for parallel MR imaging at ultrahigh fields.

Authors:  Yong Pang; Ernest W H Wong; Baiying Yu; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2014-02

Review 4.  Massively parallel MRI detector arrays.

Authors:  Boris Keil; Lawrence L Wald
Journal:  J Magn Reson       Date:  2013-02-07       Impact factor: 2.229

5.  Resonant Mode Reduction in Radiofrequency Volume Coils for Ultrahigh Field Magnetic Resonance Imaging.

Authors:  Yong Pang; Zhentian Xie; Ye Li; Duan Xu; Daniel Vigneron; Xiaoliang Zhang
Journal:  Materials (Basel)       Date:  2011-07-28       Impact factor: 3.623