A new peripherally implanted subcutaneous permanent central venous access device for patients requiring chemotherapy.

R R Salem1, B A Ward, T S Ravikumar.   

Abstract

PURPOSE: A peripherally implanted central venous access device (P.A.S. Port; Pharmacia Deltec Inc, St Paul, MN) was evaluated for ease of insertion, functionality, acceptance, and complications in patients who required long-term venous access. A hand-held tracking system (Cath-Finder; Pharmacia Deltec Inc) used to determine catheter tip location was also evaluated. PATIENTS AND METHODS: A P.A.S. Port was placed in 47 patients who required long-term intravenous access. The median follow-up duration has been 32 weeks (range, 2 to 112). Total usage has been 2,028 catheter-weeks. The Cath-Finder was used to determine catheter tip location during insertion. Nursing staff and patient satisfaction were polled and functionality and complications were recorded.
RESULTS: The device was found to be simple to insert, the procedure well tolerated, and, with one exception, the Cath-Finder accurately predicted catheter tip location. There was a 6.4% incidence of transient phlebitis and a 6.4% incidence of symptomatic axillary or subclavian vein thrombosis. There were no infectious complications. Access was simple in all but two obese patients. The device functioned well in all patients, except three in whom blood aspiration was difficult and two in whom fluid administration was slow. The device was well tolerated by all patients and nursing staff satisfaction was high.
CONCLUSION: This device provides a highly acceptable, additional method of implantable, permanent central venous access for chemotherapy patients with a low complication rate. The successful use of the Cath-Finder and minor extent of the procedure may allow this device to be inserted in a clinic procedure room without sedation and fluoroscopy.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 8229132     DOI: 10.1200/JCO.1993.11.11.2181

Source DB:  PubMed          Journal:  J Clin Oncol        ISSN: 0732-183X            Impact factor:   44.544


  8 in total

1.  Microcatheter tip enhancement in fluoroscopy: a comparison of techniques.

Authors:  Akihiro Takemura; Kenneth R Hoffmann; Masayuki Suzuki; Zhou Wang; Hussain S Rangwala; Hajime Harauchi; Stephen Rudin; Tokuo Umeda
Journal:  J Digit Imaging       Date:  2007-12       Impact factor: 4.056

2.  An algorithm for tracking microcatheters in fluoroscopy.

Authors:  Akihiro Takemura; Kenneth R Hoffmann; Masayuki Suzuki; Zhou Wang; Hussain S Rangwala; Hajime Harauchi; Stephen Rudin; Tokuo Umeda
Journal:  J Digit Imaging       Date:  2008-03       Impact factor: 4.056

3.  Current situation regarding central venous port implantation procedures and complications: a questionnaire-based survey of 11,693 implantations in Japan.

Authors:  Masatoshi Shiono; Shin Takahashi; Masanobu Takahashi; Takuhiro Yamaguchi; Chikashi Ishioka
Journal:  Int J Clin Oncol       Date:  2016-06-21       Impact factor: 3.402

4.  Upper arm central venous port implantation: a 6-year single institutional retrospective analysis and pictorial essay of procedures for insertion.

Authors:  Masatoshi Shiono; Shin Takahashi; Yuichi Kakudo; Masanobu Takahashi; Hideki Shimodaira; Shunsuke Kato; Chikashi Ishioka
Journal:  PLoS One       Date:  2014-03-10       Impact factor: 3.240

5.  A Study of Use of "PORT" Catheter in Patients with Cancer: A Single-Center Experience.

Authors:  Irappa Madabhavi; Apurva Patel; Malay Sarkar; Asha Anand; Harsha Panchal; Sonia Parikh
Journal:  Clin Med Insights Oncol       Date:  2017-02-23

6.  Arm port vs chest port: a systematic review and meta-analysis.

Authors:  Guanhua Li; Yu Zhang; Hongmin Ma; Junmeng Zheng
Journal:  Cancer Manag Res       Date:  2019-07-03       Impact factor: 3.989

Review 7.  Long-term outcome of radiological-guided insertion of implanted central venous access port devices (CVAPD) for the delivery of chemotherapy in cancer patients: institutional experience and review of the literature.

Authors:  J Vardy; K Engelhardt; K Cox; J Jacquet; A McDade; M Boyer; P Beale; M Stockler; R Loneragan; B Dennien; R Waugh; S J Clarke
Journal:  Br J Cancer       Date:  2004-09-13       Impact factor: 7.640

8.  Central venous port system associated thromboses: outcome in 3498 implantations and literature review.

Authors:  Martina Schumacher; Roland H Wagner
Journal:  Ger Med Sci       Date:  2007-09-03
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.