Literature DB >> 8145585

Experimental acquisition system for impedance tomography with active electrode approach.

B Rigaud1, Y Shi, N Chauveau, J P Morucci.   

Abstract

An experimental system for impedance tomography has been constructed. The acquisition system uses 16 multifunctional active electrodes, each including a current source and a voltage buffer. Images of active and reactive parts of different target impedances in a phantom filled with liquid have been obtained. The system performance has been compared with those of other systems using either a mesh phantom or rods as point sources used for the determination of the modulation transfer function.

Mesh:

Year:  1993        PMID: 8145585     DOI: 10.1007/bf02441807

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  8 in total

1.  Distinguishability of conductivities by electric current computed tomography.

Authors:  D Isaacson
Journal:  IEEE Trans Med Imaging       Date:  1986       Impact factor: 10.048

2.  Detection and imaging of electric conductivity and permittivity at low frequency.

Authors:  L F Fuks; M Cheney; D Isaacson; D G Gisser; J C Newell
Journal:  IEEE Trans Biomed Eng       Date:  1991-11       Impact factor: 4.538

3.  Skin impedance from 1 Hz to 1 MHz.

Authors:  J Rosell; J Colominas; P Riu; R Pallas-Areny; J G Webster
Journal:  IEEE Trans Biomed Eng       Date:  1988-08       Impact factor: 4.538

4.  A phantom for electrical impedance tomography.

Authors:  H Griffiths
Journal:  Clin Phys Physiol Meas       Date:  1988

5.  Aspects of instrumentation design for impedance imaging.

Authors:  D Murphy; P Rolfe
Journal:  Clin Phys Physiol Meas       Date:  1988

6.  Limitations in hardware design in impedance imaging.

Authors:  A D Seagar; B H Brown
Journal:  Clin Phys Physiol Meas       Date:  1987

7.  The Sheffield data collection system.

Authors:  B H Brown; A D Seagar
Journal:  Clin Phys Physiol Meas       Date:  1987

8.  Impedance imaging in the newborn.

Authors:  D Murphy; P Burton; R Coombs; L Tarassenko; P Rolfe
Journal:  Clin Phys Physiol Meas       Date:  1987
  8 in total
  1 in total

Review 1.  Front-end architecture for a multi-frequency electrical impedance tomography system.

Authors:  D Jennings; I D Schneider
Journal:  Med Biol Eng Comput       Date:  2001-05       Impact factor: 2.602

  1 in total