Aspects of instrumentation design for impedance imaging.

D Murphy1, P Rolfe.   

Abstract

This paper takes the circuit for a general purpose impedance imaging data measurement system and analyses the importance of several components. For high accuracy it is shown that stray capacitances in the input circuitry are the major problem although the use of a phase sensitive demodulator can minimise this problem. It is shown that harmonic distortion in the constant current generator can cause errors which are, however, reduced if a linear multiplying demodulator is used rather than a switching demodulator.

Mesh:

Year:  1988        PMID: 3240650     DOI: 10.1088/0143-0815/9/4a/002

Source DB:  PubMed          Journal:  Clin Phys Physiol Meas        ISSN: 0143-0815