In pulsed Doppler systems the received RF (radio frequency) signal is multiplied by a quadrature reference signal and subsequently averaged over a short depth range to obtain a sample of the complex Doppler signal. The mean frequency of the sampled Doppler signal, obtained with the autocorrelation function, reflects the mean velocity of the scatterers moving through the sample volume. An alternative is to evaluate the two-dimensional cross correlation function of a short segment of the RF signals over subsequent lines, giving the mean velocity of the scatterers. Both methods of velocity estimation were applied to computer-generated RF signals with varying RF bandwidth, signal-to-noise ratio, and mean and width of the imposed velocity distribution. The length of the RF signal segment and the number of lines for velocity estimation (package length) affects the accuracy of the velocity estimate. It can be concluded that the cross correlation technique behaves superiorly especially for a low velocity dispersion. Furthermore, the standard deviation of the velocity estimate decreases for an increasing sample volume length and package length, while the performance of the conventional Doppler technique is rather independent of the length of the sample volume. The difference between both techniques decreases for a greater package length or for signals simulating a wide velocity distribution.

Mesh:

Year:  1993        PMID: 8134974     DOI: 10.1016/0301-5629(93)90090-b

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  8 in total

1.  Modeling and phantom studies of ultrasonic wall shear rate measurements using coded pulse excitation.

Authors:  Jean K Tsou; Jie Liu; Michael F Insana
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2006-04       Impact factor: 2.725

2.  Time delay estimation using wavelet transform for pulsed-wave ultrasound.

Authors:  X L Xu; A H Tewfik; J F Greenleaf
Journal:  Ann Biomed Eng       Date:  1995 Sep-Oct       Impact factor: 3.934

Review 3.  Developments in cardiovascular ultrasound: Part 1: Signal processing and instrumentation.

Authors:  P J Fish; P R Hoskins; C Moran; W N McDicken
Journal:  Med Biol Eng Comput       Date:  1997-11       Impact factor: 2.602

4.  Measurement of nonlinear property of artery wall using remote cyclic actuation.

Authors:  Hideyuki Hasegawa; Hiroshi Kanai
Journal:  J Med Ultrason (2001)       Date:  2006-09       Impact factor: 1.314

5.  An FPGA-Based Backend System for Intravascular Photoacoustic and Ultrasound Imaging.

Authors:  Xun Wu; Jean L Sanders; Xiao Zhang; Feysel Yalcin Yamaner; Omer Oralkan
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-11-14       Impact factor: 2.725

6.  Evaluation of cervical stiffness during pregnancy using semiquantitative ultrasound elastography.

Authors:  E Hernandez-Andrade; S S Hassan; H Ahn; S J Korzeniewski; L Yeo; T Chaiworapongsa; R Romero
Journal:  Ultrasound Obstet Gynecol       Date:  2013-01-08       Impact factor: 7.299

7.  Split-spectrum amplitude-decorrelation angiography with optical coherence tomography.

Authors:  Yali Jia; Ou Tan; Jason Tokayer; Benjamin Potsaid; Yimin Wang; Jonathan J Liu; Martin F Kraus; Hrebesh Subhash; James G Fujimoto; Joachim Hornegger; David Huang
Journal:  Opt Express       Date:  2012-02-13       Impact factor: 3.894

8.  A new high-resolution spectral approach to noninvasively evaluate wall deformations in arteries.

Authors:  Ivonne Bazan; Carlos Negreira; Antonio Ramos; Javier Brum; Alfredo Ramirez
Journal:  Comput Math Methods Med       Date:  2014-02-13       Impact factor: 2.238

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.