Literature DB >> 8083085

Photogrammetric accuracy measurements of head holder systems used for fractionated radiotherapy.

M Menke1, F Hirschfeld, T Mack, O Pastyr, V Sturm, W Schlegel.   

Abstract

PURPOSE: We describe how stereo photogrammetry can be used to determine immobilization and repositioning accuracies of head holder systems used for fractionated radiotherapy of intracranial lesions. METHODS AND MATERIALS: The apparatus consists of two video cameras controlled by a personal computer and a bite block based landmark system. Position and spatial orientation of the landmarks are monitored by the cameras and processed for the real-time calculation of a target point's actual position relative to its initializing position. The target's position is assumed to be invariant with respect to the landmark system. We performed two series of 30 correlated head motion measurements on two test persons. One of the series was done with a thermoplastic device, the other one with a cast device developed for stereotactic treatment at the German Cancer Research Center. Immobilization and repositioning accuracies were determined with respect to a target point situated near the base of the skull. The repositioning accuracies were described in terms of the distributions of the mean displacements of the single motion measurements.
RESULTS: Movements of the target in the order of 0.05 mm caused by breathing could be detected with a maximum resolution in time of 12 ms. The data derived from the investigation of the two test persons indicated similar immobilization accuracies for the two devices, but the repositioning errors were larger for the thermoplastic device than for the cast device. Apart from this, we found that for the thermoplastic mask the lateral repositioning error depended on the order in which the mask was closed.
CONCLUSION: The photogrammetric apparatus is a versatile tool for accuracy measurements of head holder devices used for fractionated radiotherapy.

Entities:  

Mesh:

Year:  1994        PMID: 8083085     DOI: 10.1016/0360-3016(94)90412-x

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  11 in total

1.  Evaluation of methods for opto-electronic body surface sensing applied to patient position control in breast radiation therapy.

Authors:  G Baroni; A Troia; M Riboldi; R Orecchia; G Ferrigno; A Pedotti
Journal:  Med Biol Eng Comput       Date:  2003-11       Impact factor: 2.602

2.  Clinical results of a pilot study on stereovision-guided stereotactic radiotherapy and intensity modulated radiotherapy.

Authors:  Shidong Li; Lawrence R Kleinberg; Daniele Rigamonti; Moody D Wharam; Abdul Rashid; Juan Jackson; David Djajaputra; Shenjen He; Tunisia Creasey; Theodore L DeWeese
Journal:  Technol Cancer Res Treat       Date:  2010-12

3.  Patterns of failure after stereotactic radiotherapy of intracranial meningioma.

Authors:  Vasileios Askoxylakis; Angelika Zabel-du Bois; Wolfgang Schlegel; Jürgen Debus; Peter Huber; Stefanie Milker-Zabel
Journal:  J Neurooncol       Date:  2009-12-13       Impact factor: 4.130

4.  [Noninvasive head fixation for external irradiation of tumors of the head-neck area].

Authors:  R J Bale; R Sweeney; M Vogele; M Nevinny; T Auer; A Bluhm; W F Thumfart; P Lukas
Journal:  Strahlenther Onkol       Date:  1998-07       Impact factor: 3.621

5.  Radiochemotherapy with temozolomide as re-irradiation using high precision fractionated stereotactic radiotherapy (FSRT) in patients with recurrent gliomas.

Authors:  Stephanie E Combs; Marc Bischof; Thomas Welzel; Holger Hof; Susanne Oertel; Jürgen Debus; Daniela Schulz-Ertner
Journal:  J Neurooncol       Date:  2008-05-07       Impact factor: 4.130

6.  An immobilization and localization technique for SRT and IMRT of intracranial tumors.

Authors:  Leonid B Leybovich; Anil Sethi; Nesrin Dogan; Edward Melian; Mathew Krasin; Bahman Emami
Journal:  J Appl Clin Med Phys       Date:  2002       Impact factor: 2.102

7.  Recent advances in image-guided radiotherapy for head and neck carcinoma.

Authors:  Sameer K Nath; Daniel R Simpson; Brent S Rose; Ajay P Sandhu
Journal:  J Oncol       Date:  2009-07-29       Impact factor: 4.375

8.  Intensity Modulated Radiotherapy (IMRT) and Fractionated Stereotactic Radiotherapy (FSRT) for children with head-and-neck-rhabdomyosarcoma.

Authors:  Stephanie E Combs; Wolfgang Behnisch; Andreas E Kulozik; Peter E Huber; Jürgen Debus; Daniela Schulz-Ertner
Journal:  BMC Cancer       Date:  2007-09-13       Impact factor: 4.430

9.  Long term results after fractionated stereotactic radiotherapy (FSRT) in patients with craniopharyngioma: maximal tumor control with minimal side effects.

Authors:  Semi B Harrabi; Sebastian Adeberg; Thomas Welzel; Stefan Rieken; Daniel Habermehl; Jürgen Debus; Stephanie E Combs
Journal:  Radiat Oncol       Date:  2014-09-16       Impact factor: 3.481

10.  Improvement of therapeutic index for brain tumors with daily image guidance.

Authors:  Lisa Be Shields; James M Coons; Catherine Dedich; Maria Ragains; Kristi Scalf; Todd W Vitaz; Aaron C Spalding
Journal:  Radiat Oncol       Date:  2013-12-02       Impact factor: 3.481

View more