Fractures of the tibial plateau: value of spiral CT coronal plane reconstructions for detecting displacement in vitro.

K W McEnery1, A J Wilson, T K Pilgram, W A Murphy, M M Marushack.   

Abstract

OBJECTIVE: The purpose of this in vitro study was to determine the value of spiral CT for detecting displacement of fractures of the tibial plateau. The exact amount of inferior plateau depression, if any, is a primary criterion for deciding between surgical management and conservative management.
MATERIALS AND METHODS: An artificial fracture was produced in a cadaveric tibial plateau. Inferior displacements of 0, 1, 3, and 5 mm were created at the fracture. Spiral CT scans of each displacement were obtained with table speeds of 2, 3, and 5 mm/sec. Section collimation equaled table increment. Coronal image reconstructions were produced by use of standard scanner software. The images were interpreted by six musculoskeletal radiologists.
RESULTS: Interpretation accuracy was greatest with coronal images created from the 2 mm/sec scans. For distinguishing 5-mm fracture displacements, the average diagnostic sensitivity and specificity were 96% and 93%, respectively; when a 2-mm depression was used as the criterion for clinical significance, the sensitivity was 100% and the specificity was 69%.
CONCLUSION: When minimal table increment and collimation are used, spiral CT can detect clinically important inferior depressions of tibial plateau fractures. On the basis of the results of this study, when spiral CT is used for tibial plateau fracture assessment, we recommend 2-mm section collimation, 2-mm table speed, and reconstruction of images at 1-mm increments.

Entities:  

Mesh:

Year:  1994        PMID: 7976896     DOI: 10.2214/ajr.163.5.7976896

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  3 in total

1.  [The isocentric C-arm. Visualization of fracture reduction and screw position in the radius].

Authors:  R Meier; J Geerling; T Hüfner; M Kfuri; C Krettek
Journal:  Unfallchirurg       Date:  2011-07       Impact factor: 1.000

2.  Missed rib fractures on evaluation of initial chest CT for trauma patients: pattern analysis and diagnostic value of coronal multiplanar reconstruction images with multidetector row CT.

Authors:  S H Cho; Y M Sung; M S Kim
Journal:  Br J Radiol       Date:  2012-04-18       Impact factor: 3.039

3.  3D imaging with an isocentric mobile C-arm comparison of image quality with spiral CT.

Authors:  Dorothea Kotsianos; Stefan Wirth; Tanja Fischer; Ekkehard Euler; Clemens Rock; Ulrich Linsenmaier; Klaus Jürgen Pfeifer; Maximilian Reiser
Journal:  Eur Radiol       Date:  2004-04-01       Impact factor: 5.315

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.