A hybrid method of attenuation correction for positron emission tomography brain studies.

V Bettinardi1, M C Gilardi, S Cargnel, G Rizzo, M Teräs, G Striano, F Fazio.   

Abstract

A hybrid method for attenuation correction (HAC) in positron emission tomography (PET) brain studies is proposed. The technique requires the acquisition of two short (1 min) transmission scans immediately before or after the emission study, with the patient and the head fixation system in place and after removing the patient from the scanner with the head fixation system alone. The method combines a uniform map of attenuation coefficients for the patient's head with measured attenuation coefficients for the head fixation system to generate a hybrid attenuation map. The HAC method was calibrated on 30 PET cerebral studies for comparison with the conventional measured attenuation correction method by ROI analysis. Average differences of less than 3% were found for cortical and subcortical regions. The HAC technique is particularly suitable in a PET clinical environment, allowing a reduction of the total study time, greater comfort for patients and an increase in patient throughput.

Entities:  

Mesh:

Year:  1994        PMID: 7875164     DOI: 10.1007/bf02426690

Source DB:  PubMed          Journal:  Eur J Nucl Med        ISSN: 0340-6997


  12 in total

1.  A segmented attenuation correction for PET.

Authors:  E Z Xu; N A Mullani; K L Gould; W L Anderson
Journal:  J Nucl Med       Date:  1991-01       Impact factor: 10.057

2.  Validation of simultaneous PET emission and transmission scans.

Authors:  C J Thompson; N Ranger; A C Evans; A Gjedde
Journal:  J Nucl Med       Date:  1991-01       Impact factor: 10.057

3.  Misalignment between PET transmission and emission scans: its effect on myocardial imaging.

Authors:  M E McCord; S L Bacharach; R O Bonow; V Dilsizian; A Cuocolo; N Freedman
Journal:  J Nucl Med       Date:  1992-06       Impact factor: 10.057

4.  Measured attenuation correction methods.

Authors:  H Ostertag; W K Kübler; J Doll; W J Lorenz
Journal:  Eur J Nucl Med       Date:  1989

5.  The application of a masked orbiting transmission source for attenuation correction in PET.

Authors:  N T Ranger; C J Thompson; A C Evans
Journal:  J Nucl Med       Date:  1989-06       Impact factor: 10.057

6.  Quantitation in positron emission computed tomography: 2. Effects of inaccurate attenuation correction.

Authors:  S C Huang; E J Hoffman; M E Phelps; D E Kuhl
Journal:  J Comput Assist Tomogr       Date:  1979-12       Impact factor: 1.826

7.  A method for postinjection PET transmission measurements with a rotating source.

Authors:  R E Carson; M E Daube-Witherspoon; M V Green
Journal:  J Nucl Med       Date:  1988-09       Impact factor: 10.057

8.  Attenuation correction using count-limited transmission data in positron emission tomography.

Authors:  S R Meikle; M Dahlbom; S R Cherry
Journal:  J Nucl Med       Date:  1993-01       Impact factor: 10.057

9.  Determination of object contour from projections for attenuation correction in cranial positron emission tomography.

Authors:  M Bergström; J Litton; L Eriksson; C Bohm; G Blomqvist
Journal:  J Comput Assist Tomogr       Date:  1982-04       Impact factor: 1.826

10.  A boundary method for attenuation correction in positron computed tomography.

Authors:  S C Huang; R E Carson; M E Phelps; E J Hoffman; H R Schelbert; D E Kuhl
Journal:  J Nucl Med       Date:  1981-07       Impact factor: 10.057

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.