P wave recognition and atrial stimulation with fractally iridium coated VDD single pass leads.

J C Res1, F J Reijsoo, R J van Woersem, C J Middelhoff, P M van Kalmthout, P Zwart, P Westendorp.   

Abstract

UNLABELLED: The bottleneck of VDD systems is the reliable detection of the small atrial signals by a floating atrial electrode. Fractally iridium coated electrodes offer excellent sensing and pacing performance. In this study, the performance of such a floating atrial lead in P wave sensing and synchronous ventricular stimulation was examined. Atrial pacing was also used as a test of atrial wall contact. PATIENTS AND METHODS: A fractally iridium coated VDD lead was implanted in 18 patients. In 15 patients it was interfaced with a VDD pacemaker and in 3 patients with a DDD system depending on the P wave amplitude measured acutely (> or = 2 mV). Simultaneous recordings of the surface ECG and pacemaker telemetry were used to analyze P wave amplitudes and AV synchrony in different body positions, and during normal and deep breathing. Additionally, exercise tests based on daily life activities and 24-hour ECG monitoring were performed to test the pacemaker function.
RESULTS: During implantation P wave amplitudes were 1.86 mV +/- 1.08 mV (range 0.5-4.9 mV) and during follow-up (6.6 +/- 5.6 weeks) 0.18-3.8 mV. Holter recordings revealed reliable P wave sensing at a sensitivity setting of 0.5 mV (95.5%). P wave sensing was further improved by a higher atrial sensitivity. AV synchronous pacing > or = 99.9% was achieved in all patients. In 7 patients the atrial electrode could be positioned close to the atrial wall enabling atrial stimulation thresholds at an average of 4.3 volts.
CONCLUSION: This fractally iridium coated VVD lead allowed consistent and reliable P wave sensing at an atrial sensitivity as low as 0.5 mV in selected patients.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7845786     DOI: 10.1111/j.1540-8159.1994.tb03768.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Development of sinus node disease in patients with AV block: implications for single lead VDD pacing.

Authors:  U K Wiegand; F Bode; R Schneider; A Brandes; H Haase; H A Katus; J Potratz
Journal:  Heart       Date:  1999-06       Impact factor: 5.994

2.  Fractally coated myocardial pacemaker leads in children.

Authors:  Thomas Borth-Bruhns; Matthias Gass
Journal:  J Interv Card Electrophysiol       Date:  2005-10       Impact factor: 1.900

3.  Predictors of loss of atrioventricular synchrony in single lead VDD pacing.

Authors:  P Hunziker; P Buser; M Pfisterer; F Burkart; S Osswald
Journal:  Heart       Date:  1998-10       Impact factor: 5.994

Review 4.  The current status of single lead dual chamber sensing and pacing.

Authors:  H F Tse; C P Lau
Journal:  J Interv Card Electrophysiol       Date:  1998-09       Impact factor: 1.900

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.