Fractally coated myocardial pacemaker leads in children.

Thomas Borth-Bruhns1, Matthias Gass.   

Abstract

UNLABELLED: We report our single centre experience with a new fractally coated myocardial unipolar lead (ELC35UP; Biotronik) in 96 pediatric patients (59% male, 41% female). Congenital heart disease (CHD) was associated in 89%. The age at implantation ranged between 2 days and 19 years, median for children with CHD 7.8 years, without CHD 4.7 years. Twenty percent of the children were younger than one year at implantation. Mean follow-up was 30 months (1-57 months). We compared our findings with a steroid eluting epicardial lead (CapSure EPI 4968; Medtronic) in 46 children with comparable age and sex-distribution. We found a lead survival of 87% after 57 months in the ELC35UP group (steroid lead: 87% after 129 months). Pacing energy thresholds were equal after 12 months (median 3.0 microjoules), but the sensing characteristics of the fractally coated lead was significantly superior to the steroid eluting lead with median R waves of 7.0 mV (steroid lead: 3.5 mV) after 12 months. Children with myocardial scar tissue requiring pacemaker therapy after surgery of CHD showed no differences in sensing and pacing thresholds in comparison to children with congenital rhythm disorders. The fractally coated screw-in lead offers technical advantages concerning the subxiphoidal implantation procedure.
CONCLUSION: Fractally coated ventricular screw-in leads represent a feasible alternative to the common steroid eluting leads- especially in children requiring pacemaker therapy after surgery for CHD.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16311937     DOI: 10.1007/s10840-005-3409-9

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  19 in total

1.  Non-invasive graft monitoring after heart transplantation: rationale to reduce the number of endomyocardial biopsies.

Authors:  B Grasser; F Iberer; S Schaffellner; D Kniepeiss; G Schreier; P Kastner; K H Tscheliessnigg
Journal:  Transpl Int       Date:  2000       Impact factor: 3.782

2.  A fractally coated, 1.3 mm2 high impedance pacing electrode.

Authors:  G Fröhlig; A Bolz; J Ströbel; M Rutz; P Lawall; H Schwerdt; M Schaldach; H Schieffer
Journal:  Pacing Clin Electrophysiol       Date:  1998-06       Impact factor: 1.976

3.  Evoked epimyocardial electrogram for rejection diagnosis after heart transplantation.

Authors:  T Auer; G Schreier; K H Tscheliessnigg; H Hutten; T Allmayr; B Grasser; A Wasler; B Petutschnigg; F Iberer; M Schaldach
Journal:  Transpl Int       Date:  1996       Impact factor: 3.782

4.  Twenty years experience with pediatric pacing: epicardial and transvenous stimulation.

Authors:  J S Sachweh; J F Vazquez-Jimenez; F A Schöndube; S H Daebritz; H Dörge; E G Mühler; B J Messmer
Journal:  Eur J Cardiothorac Surg       Date:  2000-04       Impact factor: 4.191

5.  [Measurement and analysis of monophasic action potentials using fractally coated electrodes--I].

Authors:  R Fröhlich; T Wetzig; A Bolz; M Schaldach
Journal:  Biomed Tech (Berl)       Date:  1995-06       Impact factor: 1.411

6.  Steroid-eluting epicardial pacing electrodes: six year experience of pacing thresholds in a growing pediatric population.

Authors:  N G Cutler; P P Karpawich; D Cavitt; M Hakimi; H L Walters
Journal:  Pacing Clin Electrophysiol       Date:  1997-12       Impact factor: 1.976

7.  Comparison of longevity, pacing, and sensing characteristics of steroid-eluting epicardial versus conventional endocardial pacing leads in children.

Authors:  G C Beaufort-Krol; H Mulder; D Nagelkerke; T W Waterbolk; M T Bink-Boelkens
Journal:  J Thorac Cardiovasc Surg       Date:  1999-03       Impact factor: 5.209

8.  Chronic performance of steroid-eluting epicardial leads in a growing pediatric population: a 10-year comparison.

Authors:  M Silvana Horenstein; Mehdi Hakimi; Henry Walters; Peter P Karpawich
Journal:  Pacing Clin Electrophysiol       Date:  2003-07       Impact factor: 1.976

9.  P wave recognition and atrial stimulation with fractally iridium coated VDD single pass leads.

Authors:  J C Res; F J Reijsoo; R J van Woersem; C J Middelhoff; P M van Kalmthout; P Zwart; P Westendorp
Journal:  Pacing Clin Electrophysiol       Date:  1994-11       Impact factor: 1.976

10.  Pacing activity, patient and lead survival over 20 years of permanent epicardial pacing in children.

Authors:  John D R Thomson; Michael E Blackburn; Carin Van Doorn; Ann Nicholls; Kevin G Watterson
Journal:  Ann Thorac Surg       Date:  2004-04       Impact factor: 4.330

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.