Literature DB >> 31462232

Comparison of NIV-NAVA and NCPAP in facilitating extubation for very preterm infants.

Byoung Kook Lee1, Seung Han Shin2,3, Young Hwa Jung4,5, Ee-Kyung Kim4,6, Han-Suk Kim4,6.   

Abstract

BACKGROUND: Various types of noninvasive respiratory modalities that lead to successful extubation in preterm infants have been explored. We aimed to compare noninvasive neurally adjusted ventilatory assist (NIV-NAVA) and nasal continuous positive airway pressure (NCPAP) for the postextubation stabilization of preterm infants.
METHODS: This retrospective study was divided into two distinct periods, between July 2012 and June 2013 and between July 2013 and June 2014, because NIV-NAVA was applied beginning in July 2013. Preterm infants of less than 30 weeks GA who had been intubated with mechanical ventilation for longer than 24 h and were weaned to NCPAP or NIV-NAVA after extubation were enrolled. Ventilatory variables and extubation failure were compared after weaning to NCPAP or NIV-NAVA. Extubation failure was defined when infants were reintubated within 72 h of extubation.
RESULTS: There were 14 infants who were weaned to NCPAP during Period I, and 2 infants and 16 infants were weaned to NCPAP and NIV-NAVA, respectively, during Period II. At the time of extubation, there were no differences in the respiratory severity score (NIV-NAVA 1.65 vs. NCPAP 1.95), oxygen saturation index (1.70 vs. 2.09) and steroid use before extubation. Several ventilation parameters at extubation, such as the mean airway pressure, positive end-expiratory pressure, peak inspiratory pressure, and FiO2, were similar between the two groups. SpO2 and pCO2 preceding extubation were comparable. Extubation failure within 72 h after extubation was observed in 6.3% of the NIV-NAVA group and 37.5% of the NCPAP group (P = 0.041).
CONCLUSIONS: The data in the present showed promising implications for using NIV-NAVA over NCPAP to facilitate extubation.

Entities:  

Keywords:  Airway extubation; Continuous positive airway pressure; Neurally adjusted ventilator assist; Noninvasive ventilation; Ventilator weaning

Year:  2019        PMID: 31462232      PMCID: PMC6712684          DOI: 10.1186/s12887-019-1683-4

Source DB:  PubMed          Journal:  BMC Pediatr        ISSN: 1471-2431            Impact factor:   2.125


Background

Invasive mechanical ventilation (MV) is frequently required in preterm infants after birth to maintain adequate alveolar ventilation and effective gas exchange. However, tracheal intubation and MV in preterm neonates can induce ventilator-induced lung injury (VILI) and airway inflammation [1, 2]. Prolonged MV in preterm infants also increases the risk of ventilator-associated pneumonia, increasing the length of hospital stays, mortality, and neurologic impairment [3]. Therefore, noninvasive respiratory modalities have been used in preterm infants to facilitate the transition to spontaneous breathing following extubation [4-7]. Nasal continuous positive airway pressure (NCPAP) maintains functional residual capacity while improving lung compliance and oxygenation. NCPAP has been widely used in the neonatal intensive care unit (NICU) and has proven to be effective in preventing failure of extubation in preterm infants [8]. However, studies have reported that extubation failure rates ranged from 25 to 35% among preterm infants who were given NCPAP after extubation [9, 10]. Nasal intermittent positive pressure ventilation (NIPPV) augments NCPAP by superimposing ventilator inflation on NCPAP [11]. Although synchronized (SNIPPV) or nonsynchronized techniques can be used to supplement the infants’ own breathing efforts, it is likely that more effective support can be achieved with SNIPPV [12, 13]. To date, pneumatic capsules or flow sensors have been used to detect inspiration for synchronization, but some limitations in clinical practice have been reported [14-16]. Neurally adjusted ventilatory assist (NAVA) improves synchrony in patients with respiratory support by detecting the electrical activity of the diaphragm and may offer potential benefits in neonatal ventilation [17-20]. Noninvasive ventilation using NAVA as a triggering modality (NIV-NAVA) could be effective, as demonstrated in adult populations [21, 22]. To date, few studies of NIV-NAVA in preterm infants have been conducted. Patient-ventilator synchrony and effective diaphragmatic unloading were reported in preterm infants during NAVA-derived noninvasive nasal ventilation [23]. Herein, we aimed to compare NIV-NAVA and NCPAP for the postextubation stabilization of very low birth weight infants.

Methods

This study used a retrospective approach and was approved by the Institutional Review Board of Seoul National University Hospital. The study included preterm infants of less than 30 weeks gestational age (GA) who were admitted to the NICU of the Seoul National University Children’s Hospital (SNUCH) between July 2012 and June 2014 and survived more than 72 h. Infants who were on MV for longer than 24 h and were weaned to NCPAP (Infant Flow system, Viasys, Healthcare, Pennsylvania, United States) or NIV-NAVA (SERVO-I, Maquet Critical Care AB, Solna, Sweden) after extubation were eligible for the study. The size of the Edi catheter used during the study period was 6 Fr/49 cm, which could be used for extremely preterm infants [20]. There were no postmenstrual age (PMA) criteria for the use of NIV-NAVA during the study period if self-respiration was well established in the baby. Infants who had major congenital anomalies or who were intubated for longer than 6 weeks were excluded from the study. The study period was divided into two distinct periods, namely between July 2012 and June 2013 (Period I) and between July 2013 and June 2014 (Period II), because NIV-NAVA was applied at SNUCH beginning in July 2013. The respiratory severity score (RSS = mean airway pressure (cmH2O) x FiO2) and oxygen saturation index (OSI = MAP x FiO2 × 100 ÷ SpO2) were used to compare the pre-extubation respiratory conditions between the two groups [24, 25]. The RSS has been used to predict extubation readiness or the length of mechanical ventilation in preterm infants, and the OSI has been suggested to be a useful measurement to reliably assess the severity of respiratory conditions in preterm infants when the oxygen index is not available [26, 27]. During the study period, extubation was performed if the patient remained stable with a SpO2 > 90% for at least 6 h while on the following settings: mean airway pressure (MAP) ≤ 9 cmH2O, positive end expiratory pressure (PEEP) ≤ 7 cmH2O and fraction of inspired oxygen (FiO2) ≤ 40%. In infants who were mechanically ventilated for longer than 15 days, dexamethasone was administered to reduce airway edema. All infants included in the study population were treated with caffeine. A capillary blood gas analysis was performed within 1 h after extubation. Postextubation PEEP was initially set to 5~6 cmH2O both in the NCPAP and NIV-NAVA groups, and was then adjusted within a range of 4~8 cmH2O according to the clinician’s discrimination. The NAVA level was initially set to 1.0~1.5 cmH2O/μV and adjusted to obtain pCO2 < 70 mmHg. In both ventilation strategies, binasal prongs and masks were used alternatively every 24 h to minimize nasal injury. The primary outcome of the study was extubation failure within 72 h after extubation, which was defined according to a set of conditions for reintubation and the reapplication of MV [28]. Infants with severe apnea requiring positive pressure ventilation (PPV), ≥ 4 apneic episodes per hour needing moderate stimulation, FiO2 > 60%, or uncompensated respiratory acidosis (pH < 7.25) were reintubated during the study period. Backup ventilation at a rate of 30/min and pressure of 10–15 cmH2O above PEEP was applied if Edi was absent or apnea occurred for more than 5–10 s and the upper pressure limit was set to 20–25 cmH2O [23]. All statistical analyses were performed with STATA 11.0 (Stata Corp, College Station, TX, USA) using the independent t-test for continuous variables and the χ2-test and Fisher’s exact test for categorical variables. For all statistical analyses, P < 0.05 was considered statistically significant.

Results

A total of 64 infants in Period I and 51 infants in Period II who were born at less than 30 weeks of gestation and survived greater than 72 h were admitted (Fig. 1). Two infants from Period I were excluded: one infant had Beckwith-Wiedemann syndrome, and the other infant had Galen malformation of the brain. Sixteen infants in Period I and 13 infants in Period II who were never intubated or intubated less than 24 h were also excluded. After excluding infants who had been intubated for greater than 6 weeks, those who were never extubated or died before discharge, and those who were weaned to other modalities, such as heated and humidified high flow nasal cannula (HHHFNC), there were 14 infants who were weaned to NCPAP during Period I and 16 infants who were weaned to NIV-NAVA during Period II. The 2 infants who were weaned to NCPAP during Period II were categorized as the NCPAP group with the infants from Period I.
Fig. 1

Selection of the study population during the study period

Selection of the study population during the study period The GA and birth weight of the NIV-NAVA group and NCPAP group were not significantly different (27+ 1 vs. 26+ 5 weeks and 875 vs. 845 g, respectively) (Table 1). The incidence of RDS, maternal histologic chorioamnionitis and antenatal steroid use were also not significantly different between the two groups. At the time of extubation, PMA and weight exhibited no significant differences between the NIV-NAVA group and NCPAP group (30 vs. 29+ 4 weeks and 1045 vs. 1205 g, respectively) (Table 2). No differences in RSS (NIV-NAVA 1.65 vs. NCPAP 1.95), OSI (1.70 vs. 2.09) or steroid use were noted before extubation. Several ventilation parameters at extubation, such as MAP, PEEP, PIP peak inspiratory pressure (PIP), and FiO2, were similar between the two groups. SpO2 and pCO2 preceding extubation were also comparable.
Table 1

Demographics of the study population

NIV-NAVA(n = 16)NCPAP(n = 16)P value
GA (weeks)27+ 1 (26+ 5, 27+ 6)26+ 5 (25+ 4, 27+ 6)0.317
Birth weight (grams)875 (677.5, 1145)845 (700, 1030)0.777
Male11 (68.8)7 (43.8)0.143
C/S8 (50.0)7 (43.8)0.500
Multiple births12 (75.0)10 (62.5)0.352
PIH4 (25.0)1 (6.25)0.166
hCAM5 (31.3)10 (62.5)0.078
PPROM7 (43.8)6 (37.5)0.500
Antenatal steroid7 (43.8)12 (75.0)0.074
1-min AS3 (2, 5)3.5 (2, 4.5)0.802
5-min AS5.5 (4, 7)7 (6, 7)0.122
RDS14 (87.5)16 (100)0.242
PDA12 (75.0)7 (73.3)0.618

Values are presented as the median (interquartile range) or n (%)

NIV-NAVA Noninvasive neurally adjusted ventilatory assist, NCPAP Nasal continuous positive airway pressure, GA Gestational age, C/S Cesarean section, PIH Pregnancy induced hypertension, hCAM Histologic chorioamnionitis, PPROM Preterm premature rupture of membrane, AS Apgar score, RDS Respiratory distress syndrome, PDA Patent ductus arteriosus

Table 2

Clinical characteristics at the time of extubation

NIV-NAVA(n = 16)NCPAP(n = 16)P value
PMA at extubation (weeks)30 (28+ 6, 31+ 4)29+ 4 (27+ 3, 30+ 4)0.282
Weight at extubation (grams)1045 (800, 1325)1025 (905, 1190)0.651
Pre-extubation
 Ventilator duration (days)21.5 (11.5, 27)9.5 (4.5, 34.5)0.365
 Systemic steroid use7 (43.8)5 (31.3)0.358
 RSS1.65 (1.49, 2.28)1.95 (1.68, 2.32)0.317
 OSI1.70 (1.53, 2.39)2.09 (1.76, 2.51)0.274
 MAP (cmH2O)7 (7, 7.5)8 (7, 8)0.212
 PEEP (cmH2O)5 (5, 5)5 (5, 6)0.531
 PIP (cmH2O)13 (12, 14)15 (12, 16)0.180
 FiO2 (%)0.24 (0.21, 0.31)0.25 (0.21, 0.30)0.700
 pCO2 (mmHg)53.2 (45.0, 58.4)49.1 (43.7, 65.3)0.970
 SpO2 (mmHg)95.5 (94, 98.5)96 (93.5, 97)0.760

Values are presented as the median (interquartile range) or n (%)

NIV-NAVA Noninvasive neurally adjusted ventilatory assist, NCPAP Nasal continuous positive airway pressure, PMA Postmenstrual age, RSS Respiratory severity score, OSI Oxygen saturation index, MAP Mean airway pressure, PEEP Positive end-expiratory pressure, PIP Peak inspiratory pressure

Demographics of the study population Values are presented as the median (interquartile range) or n (%) NIV-NAVA Noninvasive neurally adjusted ventilatory assist, NCPAP Nasal continuous positive airway pressure, GA Gestational age, C/S Cesarean section, PIH Pregnancy induced hypertension, hCAM Histologic chorioamnionitis, PPROM Preterm premature rupture of membrane, AS Apgar score, RDS Respiratory distress syndrome, PDA Patent ductus arteriosus Clinical characteristics at the time of extubation Values are presented as the median (interquartile range) or n (%) NIV-NAVA Noninvasive neurally adjusted ventilatory assist, NCPAP Nasal continuous positive airway pressure, PMA Postmenstrual age, RSS Respiratory severity score, OSI Oxygen saturation index, MAP Mean airway pressure, PEEP Positive end-expiratory pressure, PIP Peak inspiratory pressure Extubation failure within 72 h after extubation was ascertained in 1 (6.3%) infant in the NIV-NAVA group and 6 (37.5%) infants in the NCPAP group (P = 0.041) (Table 3). One infant in the NIV-NAVA group was reintubated 11 h after extubation because of severe apnea requiring PPV. In the NCPAP group, 3 infants were reintubated before 24 h after extubation, 2 infants were reintubated 24–48 h after extubation and one infant was reintubated 70 h after extubation (Fig. 2). Three infants were reintubated because of severe apnea requiring PPV, two infants due to uncompensated respiratory acidosis (pH < 7.25) with pCO2 > 70 mmHg and one infant due to ≥4 apneic episodes per hour needing moderate stimulation. The use of other respiratory support parameters after extubation, such as PEEP and FiO2, were comparable between the NCPAP and NIV-NAVA groups with similar pCO2 and SpO2. Among those who were reintubated in the study, GA at birth was 26.4 weeks in the NIV-NAVA group and 25.9 (25.3–28.1) weeks in the NCPAP group. In the univariate logistic regression analysis, GA at extubation and the duration of invasive ventilation before extubation were not associated with reintubation (data not shown).
Table 3

Post-extubation status of the study population

NIV-NAVA(n = 16)NCPAP(n = 16)P value
PEEP (cmH2O)6 (5.5, 6)6 (5, 7)1.000
FiO2 (%)0.30 (0.27, 0.35)0.25 (0.21, 0.33)0.109
pCO2 (mmHg)48.5 (44.3, 53.6)49.7 (40.7, 62.1)0.695
SpO2 (mmHg)96 (93, 97)96.5 (94, 98)0.597
Extubation failure ≤72 h1 (6.3)6 (37.5)0.041

Values are presented as the median (interquartile range) or n (%). Post-extubation status was checked 1 h after extubation

NIV-NAVA Noninvasive neurally adjusted ventilatory assist, NCPAP Nasal continuous positive airway pressure, PMA Postmenstrual age, RSS Respiratory severity score, OSI Oxygen saturation index, MAP Mean airway pressure, PEEP Positive end-expiratory pressure, PIP Peak inspiratory pressure

Fig. 2

Kaplan-Meier estimates for extubation success by post-extubation modality

Post-extubation status of the study population Values are presented as the median (interquartile range) or n (%). Post-extubation status was checked 1 h after extubation NIV-NAVA Noninvasive neurally adjusted ventilatory assist, NCPAP Nasal continuous positive airway pressure, PMA Postmenstrual age, RSS Respiratory severity score, OSI Oxygen saturation index, MAP Mean airway pressure, PEEP Positive end-expiratory pressure, PIP Peak inspiratory pressure Kaplan-Meier estimates for extubation success by post-extubation modality No differences were noted between the two groups regarding the other clinical outcomes, including the development of moderate to severe bronchopulmonary dysplasia (BPD) (Table 4).
Table 4

Clinical outcomes of the study population

NIV-NAVA (n=16)NCPAP (n = 16)P value
Moderate to severe BPD10 (62.5)9 (60.0)0.589
NEC ≥ stage 22 (12.5)5 (33.3)0.170
Retinopathy of prematurity4 (25.0)6 (40.0)0.306
IVH ≥ grade 22 (12.5)1 (6.7)0.525
Periventricular leukomalacia1 (6.3)0 (0)0.516

Values are presented as the median (interquartile range) or n (%)

NIV-NAVA Noninvasive neurally adjusted ventilatory assist, NCPAP Nasal continuous positive airway pressure, BPD Bronchopulmonary dysplasia, NEC Necrotizing enterocolitis, ROP Retinopathy of prematurity, IVH Intraventricular hemorrhage

Clinical outcomes of the study population Values are presented as the median (interquartile range) or n (%) NIV-NAVA Noninvasive neurally adjusted ventilatory assist, NCPAP Nasal continuous positive airway pressure, BPD Bronchopulmonary dysplasia, NEC Necrotizing enterocolitis, ROP Retinopathy of prematurity, IVH Intraventricular hemorrhage

Discussion

Extubation failure is often observed in preterm infants because the chest wall and upper airway collapses easily and diaphragmatic strength is poor [29, 30]. The present study revealed that NIV-NAVA facilitated extubation better than NCPAP. Following a period of endotracheal intubation and IPPV, NCPAP is effective for preventing extubation failure in preterm infants [8]. This technique appears to improve lung function and reduce apnea and may therefore play a role in facilitating extubation in this population. However, certain populations among preterm infants who were subject to NCPAP experienced extubation failure [6, 31–33]. NIPPV augments NCPAP by delivering ventilator breaths via nasal prongs or a mask. Although it did not improve ventilation in infants who were able to maintain their own ventilation on NCPAP, in infants with a higher baseline PaCO2, ventilation was more effectively increased by NIPPV than NCPAP [34]. Severe apnea and increased PaCO2 were the most common causes of failure in infants receiving NCPAP, and NIPPV achieved a comparative reduction in extubation failure in preterm infants. A recent meta-analysis demonstrated that the incidence of extubation failure and the need for reintubation within 48 h to 1 week was reduced by NIPPV in preterm infants [12]. However, synchronization and the device used to deliver PPV may be important parameters in NIPPV [13]. NAVA has been applied in clinical practice during the last decade, but studies have rarely involved neonates, especially the preterm infant population. However, a recent study demonstrated the effectiveness and feasibility of NAVA in this population [19]. Noninvasive support via NAVA improved patient-ventilator synchrony by reducing trigger delay and the number of asynchrony events [35]. Previously, we reported that NAVA improved patient-ventilator synchrony and diaphragmatic unloading in preterm infants during noninvasive nasal ventilation compared with pressure support mode [23]. A recent physiologic study performed by Gibu et al. compared NIV-NAVA and NIPPV and demonstrated that peak inspiratory pressure and FiO2 were lowered in NIV-NAVA than in NIPPV [36]. Furthermore, both infant movement and caretaker’s work were lowered in NIV-NAVA, suggesting that NIV-NAVA was more effective than NIPPV at increasing infant comfort. Because it has excellent synchronization, NIN-NAVA could serve as a substitute for NCPAP to facilitate extubation in preterm infants. Most cases of reintubation in this study were the result of severe apnea or uncompensated hypercapnia. When compared to NCPAP, apnea and hypercapnia were more preventable in NIPPV by generating higher airway pressure to prevent obstructive apnea and triggering sigh in preterm infants [37, 38]. Although NIV-NAVA seemed to improve ventilator synchrony and diaphragmatic unloading during noninvasive ventilation compared to other NIPPV, there was no evidence that NIV-NAVA is superior to other NIPPV modalities after extubation [23, 39]. Even though there could be concerns regarding the size of the baby when using NIV-NAVA, many studies showed NIV-NAVA was feasible in extremely preterm infants [23, 39]. In the present study, NIV-NAVA was also found to be feasible in babies as small as 660 g at extubation or 700 g at birth who were successfully weaned to NIV-NAVA at PMA 28 weeks. A baby who was 500 g at birth was also successfully weaned to NIV-NAVA at 770 g. Moreover, Edi catheters can efficiently serve as a feeding tube in these babies and thus an additional feeding tube did not need to be inserted for enteral feeding. NEC was comparable in both groups and there were no intestinal perforations or air leaks after the infants were weaned to NIV-NAVA or NCPAP. Although the rates of neonatal complications are lower in noninvasive versus invasive MV, safety must be considered. Previously, it was suggested that neonates who were mechanically ventilated with either a face mask or nasal prongs had an increased risk of gastrointestinal perforations. However, recent data has shown that NIPPV does not appear to be associated with increased gastrointestinal side effects, and the risk of air leaks was lower in NIPPV than in NCPAP [40]. No differences in the development of air leaks and NEC were observed between the two groups in the present study. There are some limitations to the present study. This study was a retrospective study with a small sample size, thus making it difficult to draw robust conclusions. There also was a period of overlap when both NIV-NAVA and NCPAP were used as weaning modalities. The study population was highly selected because we analyzed only 50% of the preterm infants born at < 30 weeks of gestation who were intubated for more than 24 h and were extubated thereafter during the study period. Furthermore, the duration of ventilation seemed to be shorter in the NCPAP group, although this result was not statistically significant. While the sample size may have been too small to fully elucidate this difference, a logistic regression analysis for reintubation was performed ad hoc and showed that the duration of ventilation before extubation was not associated with reintubation (data not shown). The criteria for extubation were well-defined in our unit, and the pre-extubation conditions in both groups including the PMA at extubation, RSS, OSI and the ventilation settings were comparable in the present study. Despite these limitations, this is the first study to compare the clinical responses between NIV-NAVA and NCPAP when used to facilitate extubation in preterm infants.

Conclusions

The data in the present study were not robust enough to be conclusive due to small sample size, but showed promising implications for using NIV-NAVA over NCPAP to facilitate extubation. NIV-NAVA could be an effective modality for synchronized noninvasive ventilation following successful extubation from MV in preterm infants.
  40 in total

Review 1.  Neural control of mechanical ventilation in respiratory failure.

Authors:  C Sinderby; P Navalesi; J Beck; Y Skrobik; N Comtois; S Friberg; S B Gottfried; L Lindström
Journal:  Nat Med       Date:  1999-12       Impact factor: 53.440

2.  Elective use of nasal continuous positive airways pressure following extubation of preterm infants.

Authors:  G Dimitriou; A Greenough; V Kavvadia; B Laubscher; C Alexiou; V Pavlou; S Mantagos
Journal:  Eur J Pediatr       Date:  2000-06       Impact factor: 3.183

Review 3.  Nasal CPAP for neonates: what do we know in 2003?

Authors:  A G De Paoli; C Morley; P G Davis
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-05       Impact factor: 5.747

4.  A prospective randomized, controlled trial comparing synchronized nasal intermittent positive pressure ventilation versus nasal continuous positive airway pressure as modes of extubation.

Authors:  M N Khalaf; N Brodsky; J Hurley; V Bhandari
Journal:  Pediatrics       Date:  2001-07       Impact factor: 7.124

5.  Variables associated with the early failure of nasal CPAP in very low birth weight infants.

Authors:  Amer Ammari; Mandhir Suri; Vladana Milisavljevic; Rakesh Sahni; David Bateman; Ulana Sanocka; Carrie Ruzal-Shapiro; Jen-Tien Wung; Richard A Polin
Journal:  J Pediatr       Date:  2005-09       Impact factor: 4.406

6.  Surfactant administration by transient intubation in infants 29 to 35 weeks' gestation with respiratory distress syndrome decreases the likelihood of later mechanical ventilation: a randomized controlled trial.

Authors:  Ann Reininger; Rubia Khalak; James W Kendig; Rita M Ryan; Timothy P Stevens; Linda Reubens; Carl T D'Angio
Journal:  J Perinatol       Date:  2005-11       Impact factor: 2.521

7.  Randomized trial of continuous positive airways pressure to prevent reventilation in preterm infants.

Authors:  M Peake; P Dillon; N J Shaw
Journal:  Pediatr Pulmonol       Date:  2005-03

8.  A randomized, controlled trial comparing two different continuous positive airway pressure systems for the successful extubation of extremely low birth weight infants.

Authors:  Beatrice M Stefanescu; W Paul Murphy; Brenda J Hansell; Mamta Fuloria; Timothy M Morgan; Judy L Aschner
Journal:  Pediatrics       Date:  2003-11       Impact factor: 7.124

9.  Ventilator-associated pneumonia in extremely preterm neonates in a neonatal intensive care unit: characteristics, risk factors, and outcomes.

Authors:  Anucha Apisarnthanarak; Galit Holzmann-Pazgal; Aaron Hamvas; Margaret A Olsen; Victoria J Fraser
Journal:  Pediatrics       Date:  2003-12       Impact factor: 7.124

Review 10.  Nasal continuous positive airways pressure immediately after extubation for preventing morbidity in preterm infants.

Authors:  P G Davis; D J Henderson-Smart
Journal:  Cochrane Database Syst Rev       Date:  2003
View more
  5 in total

1.  Less invasive surfactant administration versus endotracheal surfactant instillation followed by limited peak pressure ventilation in preterm infants with respiratory distress syndrome in China: study protocol for a randomized controlled trial.

Authors:  Jiajun Zhu; Yingying Bao; Lizhong Du; Huafei Huang; Qin Lv; Yejun Jiang; Yuxuan Dai; Zhijun Chen; Jingyun Shi; Yongyan Shi; Chuangzhong Yang; Hua Mei; Hong Jiang; Yanhui Sun; Xuemei Sun
Journal:  Trials       Date:  2020-06-11       Impact factor: 2.279

Review 2.  Non-invasive Respiratory Support of the Premature Neonate: From Physics to Bench to Practice.

Authors:  Ibrahim Sammour; Sreenivas Karnati
Journal:  Front Pediatr       Date:  2020-05-08       Impact factor: 3.418

3.  Respiratory and Gastrointestinal Management of an Infant with a Birth Weight of 258 Grams.

Authors:  Ryo Itoshima; Arata Oda; Ryo Ogawa; Toshimitsu Yanagisawa; Takehiko Hiroma; Tomohiko Nakamura
Journal:  AJP Rep       Date:  2021-10-25

4.  Positive versus negative pressure during removal of endotracheal-tube on prevention of post-extubation atelectasis in ventilated neonates: A randomized controlled trial.

Authors:  Roya Farhadi; Maryam Nakhshab; Atefeh Hojjati; Mohammad Khademloo
Journal:  Ann Med Surg (Lond)       Date:  2022-04-04

Review 5.  A Review on Non-invasive Respiratory Support for Management of Respiratory Distress in Extremely Preterm Infants.

Authors:  Yuan Shi; Hemananda Muniraman; Manoj Biniwale; Rangasamy Ramanathan
Journal:  Front Pediatr       Date:  2020-05-28       Impact factor: 3.418

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.