Intraoperative reflectance oximetry in burn patients.

R L Sheridan1, K M Prelack, L M Petras, S K Szyfelbein, R G Tompkins.   

Abstract

OBJECTIVE: Transmission oximetry sites for intraoperative monitoring are frequently difficult to find in burn patients, as standard transmission oximetry sites are often burned or contained within the operative field. The objective of this study was to determine if reflectance oximetry is of potential value in monitoring this group of patients.
METHODS: A total of 16 operative procedures in a group of acutely burned adult and pediatric patients with an average age of 9.7 years (range, 10 months to 37 years), average burn size of 42% of the body surface (range, 15% to 94%), and average weight of 34.2 kg (range, 9 to 100 kg) were done with simultaneous transmission and reflectance oximetry monitoring.
RESULTS: During these 16 procedures in a diverse group of acutely burned adult and pediatric patients, there was no significant difference in saturations derived from transmission and reflectance oximetry probes. In smaller children, adequate signal for reflectance probe monitoring was often detected in hyperemic sites, such as healed partial thickness burn.
CONCLUSIONS: This is the first published report documenting both the clinical use of the reflectance oximetry in burn patients and the clinical use of the Nellcor Oxisensor II RS-10 reflectance oximetry probe (Nellcor Corporation, Hayward, CA). This technique can facilitate the intraoperative monitoring of acutely burned adult and pediatric patients in whom standard transmission oximetry sites are difficult to find.

Entities:  

Mesh:

Year:  1995        PMID: 7745451     DOI: 10.1007/BF01627417

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  9 in total

Review 1.  Pulse oximetry. Uses and abuses.

Authors:  L M Schnapp; N H Cohen
Journal:  Chest       Date:  1990-11       Impact factor: 9.410

Review 2.  Recent developments in pulse oximetry.

Authors:  J W Severinghaus; J F Kelleher
Journal:  Anesthesiology       Date:  1992-06       Impact factor: 7.892

3.  Guidelines for standards of care for patients with acute respiratory failure on mechanical ventilatory support. Task Force on Guidelines; Society of Critical Care Medicine.

Authors: 
Journal:  Crit Care Med       Date:  1991-02       Impact factor: 7.598

4.  Comparison of reflection and transmission pulse oximetry after open-heart surgery.

Authors:  H Pälve
Journal:  Crit Care Med       Date:  1992-01       Impact factor: 7.598

5.  Design and evaluation of a new reflectance pulse oximeter sensor.

Authors:  Y Mendelson; J C Kent; B L Yocum; M J Birle
Journal:  Med Instrum       Date:  1988-08

6.  Forehead pulse oximetry compared with finger pulse oximetry and arterial blood gas measurement.

Authors:  E Y Cheng; M B Hopwood; J Kay
Journal:  J Clin Monit       Date:  1988-07

7.  Pulse oximetry.

Authors:  W New
Journal:  J Clin Monit       Date:  1985-04

8.  Pulse oximetry in critically ill children.

Authors:  C D Fait; R C Wetzel; J M Dean; C L Schleien; F R Gioia
Journal:  J Clin Monit       Date:  1985-10

9.  Pulse oximetry.

Authors:  R M Kulick
Journal:  Pediatr Emerg Care       Date:  1987-06       Impact factor: 1.454

  9 in total
  2 in total

1.  Comparison of a prototype esophageal oximetry probe with two conventional digital pulse oximetry monitors in aortocoronary bypass patients.

Authors:  R C Prielipp; P E Scuderi; M H Hines; J L Atlee; J F Butterworth
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

Review 2.  Acute and perioperative care of the burn-injured patient.

Authors:  Edward A Bittner; Erik Shank; Lee Woodson; J A Jeevendra Martyn
Journal:  Anesthesiology       Date:  2015-02       Impact factor: 7.892

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.