Corrected formula for estimating peripheral blood flow by impedance plethysmography.

G D Jindal1, S N Nerurkar, S A Pednekar, J P Babu, A C Bhuta, A K Deshpande.   

Abstract

In this paper, the limitations of the conventional formula for the computation of peripheral blood flow from impedance plethysmograms are highlighted, and a correction to the formula is suggested. A conductivity cell experiment is described to show the dependence of the value of the blood flow index (BFI), obtained from the conventional formula, on the mean resistivity of the cell. It is also shown that the value of the corrected BFI is independent of the mean resistivity. Anomalies observed in the amplitude of systolic waves in impedance plethysmograms of patients with oedema are explained.

Entities:  

Mesh:

Year:  1994        PMID: 7723420     DOI: 10.1007/bf02524237

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  13 in total

1.  Regional pulse volume and perfusion flow measurement. Electrical impedance plethysmography.

Authors:  J NYBOER
Journal:  Arch Intern Med       Date:  1960-02

2.  Diagnosis of venous disorders using impedance plethysmography.

Authors:  G D Jindal; S A Pedhnekar; S N Nerurkar; K L Masand; D K Gupta; H L Deshmukh; J P Bapu; G B Parulkar
Journal:  J Postgrad Med       Date:  1990-07       Impact factor: 1.476

3.  Diagnosis of peripheral arterial occlusive diseases using impedance plethysmography.

Authors:  G D Jindal; S N Nerurkar; S A Pedhnekar; J P Babu; M D Kelkar; A K Deshpande; G B Parulkar
Journal:  J Postgrad Med       Date:  1990-07       Impact factor: 1.476

4.  Technical aspects of impedance plethysmography.

Authors:  A C Bhuta; J P Babu; G D Jindal; G B Parulkar
Journal:  J Postgrad Med       Date:  1990-04       Impact factor: 1.476

5.  Principles of the impedance technique.

Authors:  L E Baker
Journal:  IEEE Eng Med Biol Mag       Date:  1989

6.  Impedance plethysmography for blood flow measurements in human limbs. Part 2. Influence of limb cross-sectional area.

Authors:  Y Yamamoto; T Yamamoto; P A Oberg
Journal:  Med Biol Eng Comput       Date:  1992-09       Impact factor: 2.602

Review 7.  Biomedical signal processing (in four parts). Part 1. Time-domain methods.

Authors:  R E Challis; R I Kitney
Journal:  Med Biol Eng Comput       Date:  1990-11       Impact factor: 2.602

8.  Comparison of volume changes in the forearm assessed by impedance and water-displacement plethysmography.

Authors:  J S Fleming; T K Hames; J Smallwood
Journal:  Med Biol Eng Comput       Date:  1986-07       Impact factor: 2.602

9.  Evaluation of the parallel conductor theory for measuring human limb blood flow by electrical admittance plethysmography.

Authors:  H Shimazu; K Yamakoshi; T Togawa; M Fukuoka; H Ito
Journal:  IEEE Trans Biomed Eng       Date:  1982-01       Impact factor: 4.538

10.  Calibrated impedance plethysmograph.

Authors:  M R Yablonski; J M Van De Water; B E Mount; E D Laska; R B Indech
Journal:  Am J Physiol       Date:  1980-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.