Diagnosis of peripheral arterial occlusive diseases using impedance plethysmography.

G D Jindal1, S N Nerurkar, S A Pedhnekar, J P Babu, M D Kelkar, A K Deshpande, G B Parulkar.   

Abstract

Impedance plethysmographic observations have been compared with arteriographic findings in 216 patients with peripheral arterial occlusive diseases. Impedance plethysmographic diagnosis in these patients was obtained by Parulkar's method without apriori knowledge of arteriographic diagnosis. But for minor discrepancy in the anatomical location of the block in few patients, impedance plethysmographic observations correlated very well with arteriographic findings. Impedance plethysmographic diagnosis was found to be correct in 312, wrong in 53, false negative in 8 limbs respectively, yielding a sensitivity of 97.5% and specificity of 98.1% of this technique.

Entities:  

Mesh:

Year:  1990        PMID: 2102915

Source DB:  PubMed          Journal:  J Postgrad Med        ISSN: 0022-3859            Impact factor:   1.476


  2 in total

1.  Validity and reproducibility of electrical impedance tomography for measurement of calf blood flow in healthy subjects.

Authors:  A Vonk Noordegraaf; P W Kunst; A Janse; R A Smulders; R M Heethaar; P E Postmus; T J Faes; P M de Vries
Journal:  Med Biol Eng Comput       Date:  1997-03       Impact factor: 2.602

2.  Corrected formula for estimating peripheral blood flow by impedance plethysmography.

Authors:  G D Jindal; S N Nerurkar; S A Pednekar; J P Babu; A C Bhuta; A K Deshpande
Journal:  Med Biol Eng Comput       Date:  1994-11       Impact factor: 2.602

  2 in total