A new device for measuring and recording the forces applied during laryngoscopy.

E P McCoy1, B A Austin, R K Mirakhur, K C Wong.   

Abstract

In characterising the forces exerted during laryngoscopy it is accepted that the significant force component is that which is parallel to the axis of the laryngoscope handle. This paper describes a new method of evaluating the forces exerted at laryngoscopy. A laryngoscope handle has been redesigned, incorporating a force-displacement transducer on the handle at the end opposite to where the blade is attached. The device is designed specifically to sense the axial component of force. The blade attachment block has been detached from the sleeve and connected to a steel shaft which forms the new battery compartment. This allows the axial force exerted at the blade during laryngoscopy to be transmitted along this shaft to the sensing transducer. Linear ball bearings have been introduced between the new shaft and the handle sleeve (outer casing). The rolling friction has been reduced to less than 0.1% by diamond-lapping the precision ground surface-hardened rod to produce a mirror finish. Thus, the force transmitted to the transducer is essentially constant regardless of where the load is applied along the blade. The output from the system is directly downloaded to a laptop computer and the data analysed almost instantaneously to determine the duration of laryngoscopy, the peak forces applied, the mean force with its standard deviation and graphic display of the data. Provision has been made for data entry checks, recording patient details and study data, and creating a data base for the storage and retrieval of the study details.

Entities:  

Mesh:

Year:  1995        PMID: 7710025     DOI: 10.1111/j.1365-2044.1995.tb15097.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  4 in total

1.  [Intubation-linked dental injuries. Relevance of individually adaptable tooth protection models].

Authors:  E Monaca; N Fock; M Doehn; M Winterhalter; F Wappler
Journal:  Anaesthesist       Date:  2010-04       Impact factor: 1.041

2.  Design, development, and face validation of an intubation simulation device using real-time force data feedback.

Authors:  Gopikrishna M Rao; Vishwanatha M Rao; Jeremy Juang; Justin Benoit; Allen L Feng; Phillip C Song
Journal:  Laryngoscope Investig Otolaryngol       Date:  2022-09-15

3.  Intubation biomechanics: laryngoscope force and cervical spine motion during intubation with Macintosh and Airtraq laryngoscopes.

Authors:  Bradley J Hindman; Brandon G Santoni; Christian M Puttlitz; Robert P From; Michael M Todd
Journal:  Anesthesiology       Date:  2014-08       Impact factor: 7.892

4.  Sex-Specific Intubation Biomechanics: Intubation Forces Are Greater in Male Than in Female Patients, Independent of Body Weight.

Authors:  Bradley J Hindman; Franklin Dexter; Benjamin C Gadomski; Martin J Bucx
Journal:  Cureus       Date:  2020-06-21
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.