| Literature DB >> 7670156 |
T A Shaler1, Y Tan, J N Wickham, K J Wu, C H Becker.
Abstract
The products from base-specific, dideoxy-nucleotide chain-termination DNA sequencing reactions catalyzed by the modified T7 DNA polymerase have been analyzed by using the technique of matrix-assisted laser desorption/ionization (MALDI) time-of-flight mass spectrometry. Preliminary experiments were performed to determine detection limits for a synthetic mixture of mixed-base single-stranded DNA which contained a 14-mer, a 21-mer, and a 41-mer; acceptable spectra, showing peaks for each component, were obtainable for samples that contained as little as 5 fmol per component. Initial sequencing reactions were therefore carried out on 2-pmol amounts of a short synthetic template that was 45 nucleotides in length, employing 2 pmol of 12-mer as the primer strand. This provided readable sequence information out to the 19th base past the primer. Using a 21-mer primer, nearly the entire sequence of the template could be read.Entities:
Mesh:
Substances:
Year: 1995 PMID: 7670156 DOI: 10.1002/rcm.1290091015
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419