Failure of the Ring polyethylene uncemented acetabular cup.

G D Harper1, T Bull, A G Cobb, G Bentley.   

Abstract

We report a high rate of failure of the Ring polyethylene cementless cup caused largely by granulomatous osteolysis. We have reviewed 126 prostheses inserted from 1986 to 1992 at from 11 to 90 months after surgery. There was radiological evidence of osteolytic granulomas adjacent to the external surface of the cup in 32%, appearing on average at three years from operation. In a subgroup of 59 prostheses followed for at least four years the incidence of such changes was 54%. A total of 27 cups (22%) have required revision, 21 for granulomatous loosening at an average follow-up of five years. In the retrieved prostheses there was obvious polyethylene abrasion and histological examination confirmed the presence of polyethylene wear debris. We found no significant correlation of osteolysis with cup size, although smaller cups were predominant among those having revision.

Entities:  

Mesh:

Year:  1995        PMID: 7615597

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  4 in total

1.  A combined XPS-SEM/EDX investigation on explanted UHMW polyethylene acetabular cups: possible role of silicon traces in the wear debris.

Authors:  E De Giglio; A Motta; L Quagliarella; L Sabbatini; G Solarino; P G Zambonin
Journal:  J Mater Sci Mater Med       Date:  2001-01       Impact factor: 3.896

2.  A rare case of pseudotumor formation associated with methyl methacrylate hypersensitivity in a patient following cemented total knee arthroplasty.

Authors:  Shachar Kenan; Leonard Kahn; Noga Haramati; Samuel Kenan
Journal:  Skeletal Radiol       Date:  2016-03-29       Impact factor: 2.199

3.  Pseudotumour Complicated by Implant Loosening One Year After revision Ceramic on Metal Total Hip Arthroplasty: A Case Report.

Authors:  Devendra Lakhotia; Prashant Kumar
Journal:  J Orthop Case Rep       Date:  2017 Jan-Feb

4.  Pseudotumor and Subsequent Implant Loosening as a Complication of Revision Total Hip Arthroplasty with Ceramic-on-Metal Bearing: A Case Report.

Authors:  Lokesh Gudda Naik; Won Yong Shon; I C Clarke; Jun-Gyu Moon; Piyush Mukund; Sang-Min Kim
Journal:  Hip Pelvis       Date:  2018-12-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.