Literature DB >> 7610122

An investigation into the comparison between different dosimetric methods of measuring profiles and depth doses for dynamic wedges on a Varian 600C linear accelerator.

P J Elder1, F M Coveney, A D Welsh.   

Abstract

Three dosimetric methods for commissioning the dynamic wedges on a Varian 600C linear accelerator operating at 4.7 MV were compared. The techniques involved the use of ionization chambers, x-ray verification film and solid state detectors. Ionization chambers gave the most accurate results, as expected, but data acquisition was very time consuming. Film dosimetry was found to have good spatial resolution for beam profiles, but problems were encountered with the acquisition of depth doses. Solid state detectors generally showed reasonable agreement with the ionization chambers.

Entities:  

Mesh:

Year:  1995        PMID: 7610122     DOI: 10.1088/0031-9155/40/4/013

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Dosimetric aspects of physical and dynamic wedge of Clinac 2100C linear accelerator.

Authors:  J S Avadhani; A S Pradhan; A Sankar; P S Viswanathan
Journal:  Strahlenther Onkol       Date:  1997-10       Impact factor: 3.621

2.  Comparison of dosimetric characteristics of physical wedge and enhanced dynamic wedge in inhomogeneous medium using Monte Carlo simulations.

Authors:  Seied Rabi Mahdavi; Atefeh Mahmoudi; Ghazale Geraily; Ahmad Mostaar; Golbarg Esmaili
Journal:  Rep Pract Oncol Radiother       Date:  2021-02-25

3.  FBX aqueous chemical dosimeter for measurement of virtual wedge profiles.

Authors:  Manoj K Semwal; Anil K Bansal; Pradeep K Thakur; Pandit B Vidyasagar
Journal:  J Appl Clin Med Phys       Date:  2008-10-24       Impact factor: 2.102

  3 in total