An image correlation procedure for digitally reconstructed radiographs and electronic portal images.

L Dong1, A L Boyer.   

Abstract

PURPOSE: To study a procedure that uses megavoltage digitally reconstructed radiographs (DRRs) calculated from patient's three-dimensional (3D) computed tomography (CT) data as a reference image for correlation with on-line electronic portal images (EPIs) to detect patient setup errors. METHODS AND MATERIALS: Megavoltage DRRs were generated by ray tracing through a modified volumetric CT data set in which CT numbers were converted into linear attenuation coefficients for the therapeutic beam energy. The DRR transmission image was transformed to the grayscale window of the EPI by a histogram-matching technique. An alternative approach was to calibrate the transmission DRR using a measured response curve of the electronic portal imaging device (EPID). This forces the calculated transmission fluence values to be distributed in the same range as that of the EPID image. A cross-correlation technique was used to determine the degree of alignment of the patient anatomy found in the EPID image relative to the reference DRR.
RESULTS: Phantom studies demonstrated that the correlation procedure had a standard deviation of 0.5 mm and 0.5 degrees in aligning translational shifts and in-plane rotations. Systematic errors were found between a reference DRR and a reference EPID image. The automated grayscale image-correlation process was completed within 3 s on a workstation computer or 12 s on a PC.
CONCLUSION: The alignment procedure allows the direct comparison of a patient's treatment portal designed with a 3D planning computer with a patient's on-line portal image acquired at the treatment unit. The image registration process is automated to the extent that it requires minimal user intervention, and it is fast and accurate enough for on-line clinical applications.

Entities:  

Mesh:

Year:  1995        PMID: 7493831     DOI: 10.1016/0360-3016(95)02082-9

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  10 in total

1.  A multistage registration method using texture features.

Authors:  Andreja Jarc; Janez Pers; Stanislav Kovacic
Journal:  J Digit Imaging       Date:  2009-01-29       Impact factor: 4.056

2.  Quality assurance for kilo- and megavoltage in-room imaging and localization for off- and online setup error correction.

Authors:  James M Balter; Larry E Antonuk
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

3.  A Spatial Registration Toolbox for Structural MR Imaging of the Aging Brain.

Authors:  Marco Ganzetti; Quanying Liu; Dante Mantini
Journal:  Neuroinformatics       Date:  2018-04

4.  A gradient feature weighted Minimax algorithm for registration of multiple portal images to 3DCT volumes in prostate radiotherapy.

Authors:  Sudhakar Chelikani; Kailasnath Purushothaman; Jonathan Knisely; Zhe Chen; Ravinder Nath; Ravi Bansal; James Duncan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-06-01       Impact factor: 7.038

5.  Accuracy of an automatic patient-positioning system based on the correlation of two edge images in radiotherapy.

Authors:  Myonggeun Yoon; Minho Cheong; Jinsung Kim; Dong Ho Shin; Sung Yong Park; Se Byeong Lee
Journal:  J Digit Imaging       Date:  2010-02-02       Impact factor: 4.056

6.  Guide to clinical use of electronic portal imaging.

Authors:  M G Herman; J J Kruse; C R Hagness
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

7.  Detection of patient setup errors with a portal image - DRR registration software application.

Authors:  Kenneth Sutherland; Masayori Ishikawa; Gerard Bengua; Yoichi M Ito; Yoshiko Miyamoto; Hiroki Shirato
Journal:  J Appl Clin Med Phys       Date:  2011-02-18       Impact factor: 2.102

8.  Definition of internal target volumes based on planar X-ray fluoroscopic images for lung and hepatic stereotactic body radiation therapy. Comparison to inhale/exhale CT technique.

Authors:  David Sevillano; Luis Miguel Núñez; Margarita Chevalier; Feliciano García-Vicente
Journal:  J Appl Clin Med Phys       Date:  2020-05-30       Impact factor: 2.102

9.  ESTERR-PRO: a setup verification software system using electronic portal imaging.

Authors:  Pantelis A Asvestas; Konstantinos K Delibasis; Nikolaos A Mouravliansky; George K Matsopoulos
Journal:  Int J Biomed Imaging       Date:  2007

10.  DRRGenerator: A Three-dimensional Slicer Extension for the Rapid and Easy Development of Digitally Reconstructed Radiographs.

Authors:  Lance Levine; Marc Levine
Journal:  J Clin Imaging Sci       Date:  2020-10-29
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.