Accuracy of an automatic patient-positioning system based on the correlation of two edge images in radiotherapy.

Myonggeun Yoon1, Minho Cheong, Jinsung Kim, Dong Ho Shin, Sung Yong Park, Se Byeong Lee.   

Abstract

We have clinically evaluated the accuracy of an automatic patient-positioning system based on the image correlation of two edge images in radiotherapy. Ninety-six head & neck images from eight patients undergoing proton therapy were compared with a digitally reconstructed radiograph (DRR) of planning CT. Two edge images, a reference image and a test image, were extracted by applying a Canny edge detector algorithm to a DRR and a 2D X-ray image, respectively, of each patient before positioning. In a simulation using a humanoid phantom, performed to verify the effectiveness of the proposed method, no registration errors were observed for given ranges of rotation, pitch, and translation in the x, y, and z directions. For real patients, however, there were discrepancies between the automatic positioning method and manual positioning by physicians or technicians. Using edged head coronal- and sagittal-view images, the average differences in registration between these two methods for the x, y, and z directions were 0.11 cm, 0.09 cm and 0.11 cm, respectively, whereas the maximum discrepancies were 0.34 cm, 0.38 cm, and 0.50 cm, respectively. For rotation and pitch, the average registration errors were 0.95° and 1.00°, respectively, and the maximum errors were 3.6° and 2.3°, respectively. The proposed automatic patient-positioning system based on edge image comparison was relatively accurate for head and neck patients. However, image deformation during treatment may render the automatic method less accurate, since the test image many differ significantly from the reference image.

Entities:  

Mesh:

Year:  2010        PMID: 20127267      PMCID: PMC3056963          DOI: 10.1007/s10278-009-9269-6

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  21 in total

1.  Automatic on-line electronic portal image analysis with a wavelet-based edge detector.

Authors:  O Petrascu; A Bel; N Linthout; D Verellen; G Soete; G Storme
Journal:  Med Phys       Date:  2000-02       Impact factor: 4.071

2.  Patient set-up using portal images: 2D/2D image registration using mutual information.

Authors:  D Plattard; M Soret; J Troccaz; P Vassal; J Y Giraud; G Champleboux; X Artignan; M Bolla
Journal:  Comput Aided Surg       Date:  2000

3.  Methods for fully automated verification of patient set-up in external beam radiotherapy with polygon shaped fields.

Authors:  K Eilertsen; A Skretting; T L Tennvassås
Journal:  Phys Med Biol       Date:  1994-06       Impact factor: 3.609

4.  A computational approach to edge detection.

Authors:  J Canny
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  1986-06       Impact factor: 6.226

5.  Digital step edges from zero crossing of second directional derivatives.

Authors:  R M Haralick
Journal:  IEEE Trans Pattern Anal Mach Intell       Date:  1984-01       Impact factor: 6.226

6.  Optimization of beam weights under dose-volume restrictions.

Authors:  M Langer; J Leong
Journal:  Int J Radiat Oncol Biol Phys       Date:  1987-08       Impact factor: 7.038

7.  An image correlation procedure for digitally reconstructed radiographs and electronic portal images.

Authors:  L Dong; A L Boyer
Journal:  Int J Radiat Oncol Biol Phys       Date:  1995-12-01       Impact factor: 7.038

8.  An image registration scheme applied to verification of radiation therapy.

Authors:  K W Leszczynski; S Loose; S Boyko
Journal:  Br J Radiol       Date:  1998-04       Impact factor: 3.039

9.  Compensating for heterogeneities in proton radiation therapy.

Authors:  M Urie; M Goitein; M Wagner
Journal:  Phys Med Biol       Date:  1984-05       Impact factor: 3.609

10.  A dose-volume-based tool for evaluating and ranking IMRT treatment plans.

Authors:  Moyed M Miften; Shiva K Das; Min Su; Lawrence B Marks
Journal:  J Appl Clin Med Phys       Date:  2004-10-01       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.