Evaluation of the Sensititre system for quantitative antimicrobial drug susceptibility testing: a collaborative study.

T L Gavan, R N Jones, A L Barry.   

Abstract

This three-center collaborative study was conducted to evaluate samples of Sensititre antimicrobial microdilution panels (GIBCO/INVENEX). Sensititre minimum inhibitory concentrations of 27 bacterial isolates were compared with those obtained by a reference microdilution method. The Sensititre and microdilution minimum inhibitory concentrations were equivalent within +/- 1 dilution in 87.6% of the comparable test results. Intralaboratory reproducibilities of the Sensititre and microdilution endpoints were equivalent with 80.4 and 82.4%, respetively, of on-scale endpoints in absolute agreement. Sensititre was more reproducible among laboratories, with nearly a 10% greater agreement of triplicate results. The Sensititre microdilution test as evaluated gave results which were essentially equivalent to those obtained with a standardized microdilution method.

Mesh:

Substances:

Year:  1980        PMID: 7425606      PMCID: PMC283810          DOI: 10.1128/AAC.17.3.464

Source DB:  PubMed          Journal:  Antimicrob Agents Chemother        ISSN: 0066-4804            Impact factor:   5.191


  4 in total

1.  Antibiotic sensitivity testing. Report of an international collaborative study.

Authors:  H M Ericsson; J C Sherris
Journal:  Acta Pathol Microbiol Scand B Microbiol Immunol       Date:  1971

2.  A microdilution method for antibiotic susceptibility testing: an evaluation.

Authors:  T L Gavan; M A Town
Journal:  Am J Clin Pathol       Date:  1970-06       Impact factor: 2.493

3.  Determination of antibiotic sensitivities by the Sensititre system.

Authors:  I Phillips; C Warren; P M Waterworth
Journal:  J Clin Pathol       Date:  1978-06       Impact factor: 3.411

4.  Evaluation of the micro-media system for quantitative antimicrobial drug susceptibility testing: a collaborative study.

Authors:  A L Barry; R N Jones; T L Gavan
Journal:  Antimicrob Agents Chemother       Date:  1978-01       Impact factor: 5.191

  4 in total
  22 in total

1.  Diagnostic accuracy assessment of Sensititre and agar disk diffusion for determining antimicrobial resistance profiles of bovine clinical mastitis pathogens.

Authors:  V Saini; R G M Olde Riekerink; J T McClure; H W Barkema
Journal:  J Clin Microbiol       Date:  2011-01-26       Impact factor: 5.948

2.  Validation of the automated reading and incubation system with Sensititre plates for antimicrobial susceptibility testing.

Authors:  Kimberle C Chapin; Michael C Musgnug
Journal:  J Clin Microbiol       Date:  2003-05       Impact factor: 5.948

3.  Evaluation of Sensititre automated reading and incubation system for automated reading of Sensititre broth microdilution susceptibility plates.

Authors:  Kimberle C Chapin; Michael C Musgnug
Journal:  J Clin Microbiol       Date:  2004-02       Impact factor: 5.948

4.  Collaborative clinical laboratory evaluation of an individual MIC strip system.

Authors:  H D Isenberg; R F D'Amato; G A McKinley; L H Green; F P Canonica; L M Fredericks; J M Hochstrasser; E R Giugliano; P Alperstein; K France
Journal:  J Clin Microbiol       Date:  1987-12       Impact factor: 5.948

5.  Comparison of BIOGRAM and commercial microdilution antimicrobial susceptibility test systems.

Authors:  R L Sautter; G A Denys
Journal:  J Clin Microbiol       Date:  1987-02       Impact factor: 5.948

6.  Evaluation of the Precept microdilution MIC system for single-drug testing in individual trays.

Authors:  R G Lairscey; M T Kelly
Journal:  J Clin Microbiol       Date:  1984-03       Impact factor: 5.948

7.  A comparison of the Api-10M commercial microdilution system with the tube dilution and standard microdilution methods.

Authors:  M Papapetropoulou; A Pappa; H Giamarellou
Journal:  Infection       Date:  1984 Mar-Apr       Impact factor: 3.553

8.  Evaluation of the Sceptor microdilution antibiotic susceptibility testing system: a collaborative investigation.

Authors:  R N Jones; C Thornsberry; A L Barry; T L Gavan
Journal:  J Clin Microbiol       Date:  1981-01       Impact factor: 5.948

9.  Comparison of sensititre dried microtitration trays with a standard agar method for determination of minimum inhibitory concentrations of antimicrobial agents.

Authors:  D S Reeves; A Holt; M J Bywater; R Wise; M N Logan; J M Andrews; J M Broughall
Journal:  Antimicrob Agents Chemother       Date:  1980-12       Impact factor: 5.191