Literature DB >> 12734233

Validation of the automated reading and incubation system with Sensititre plates for antimicrobial susceptibility testing.

Kimberle C Chapin1, Michael C Musgnug.   

Abstract

The present study compared the antimicrobial susceptibility testing (AST) results generated by the Automated Incubation and Reading System (ARIS) with custom Sensititre plates (TREK Diagnostic Systems, Cleveland, Ohio) and MicroScan PC10 GP and NUMIC10 GN plates interpreted with the WalkAway-96 system (Dade Behring, West Sacramento, Calif.) for gram-positive (GP) and gram-negative (GN) organisms as part of an in-house validation. A total of 326 isolates (3,689 antimicrobial agent-organism combinations) were evaluated. Sensititre plates were inoculated according to the instructions of the manufacturer with a suspension adjusted to a 0.5 McFarland standard, while the Prompt Inoculation System was used for the MicroScan plates. ARIS and the WalkAway system were used for automated reading of the Sensititre and MicroScan plates, respectively, at 18 to 24 h. The results were analyzed for essential (+/-1 twofold dilution) and categorical (sensitive, intermediate, or resistant) agreements. Plates that resulted in ARIS interpretations with major (falsely resistant) or very major (falsely susceptible) errors compared to the results obtained with the WalkAway system were read manually to corroborate instrument readings. Isolates for which very major or major errors were obtained and for which the results were not resolved by manual reading were retested in parallel. Isolates for which very major or major errors were obtained and for which the results were not resolved upon repeat testing were tested by the National Committee for Clinical Laboratory Standards M7-A5 frozen reference microdilution method. Essential agreement was 95.8% for 246 GN isolates. The following categorical error rates were obtained for the GN isolates: 1.3% minor errors, 0% major errors, and 0.4% very major errors. For 95 GP isolates, there was 93.5% essential agreement. Categorical error rates for GP isolates were 0.9% minor errors, 0.6% major errors, and 0.4% very major errors. ARIS-Sensititre is a diagnostic system feasible for use for automated AST in a clinical laboratory.

Entities:  

Mesh:

Year:  2003        PMID: 12734233      PMCID: PMC154729          DOI: 10.1128/JCM.41.5.1951-1956.2003

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  14 in total

1.  Imipenem susceptibility testing with a commercially prepared dry-format microdilution tray.

Authors:  J L Staneck
Journal:  J Clin Microbiol       Date:  1986-06       Impact factor: 5.948

2.  Rapid and overnight microdilution antibiotic susceptibility testing with the Sensititre Breakpoint Autoreader system.

Authors:  F S Nolte; K K Krisher; L A Beltran; N P Christianson; G E Sheridan
Journal:  J Clin Microbiol       Date:  1988-06       Impact factor: 5.948

3.  Influence of inoculum growth phase on microdilution susceptibility tests.

Authors:  A L Barry; R E Badal; R W Hawkinson
Journal:  J Clin Microbiol       Date:  1983-09       Impact factor: 5.948

4.  Evaluation of the Prompt inoculation system for preparation of standardized bacterial inocula.

Authors:  M E Lund; R W Hawkinson
Journal:  J Clin Microbiol       Date:  1983-07       Impact factor: 5.948

5.  Concurrent comparability of automated systems and commercially prepared microdilution trays for susceptibility testing.

Authors:  S L Hansen; P K Freedy
Journal:  J Clin Microbiol       Date:  1983-05       Impact factor: 5.948

6.  Automated reading of MIC microdilution trays containing fluorogenic enzyme substrates with the Sensititre Autoreader.

Authors:  J L Staneck; S D Allen; E E Harris; R C Tilton
Journal:  J Clin Microbiol       Date:  1985-08       Impact factor: 5.948

7.  Evaluation of the sensititre microdilution antibiotic susceptibility system against recent clinical isolates: three-laboratory collaborative study.

Authors:  R N Jones; T L Gavan; A L Barry
Journal:  J Clin Microbiol       Date:  1980-04       Impact factor: 5.948

8.  Evaluation of the Sensititre system for quantitative antimicrobial drug susceptibility testing: a collaborative study.

Authors:  T L Gavan; R N Jones; A L Barry
Journal:  Antimicrob Agents Chemother       Date:  1980-03       Impact factor: 5.191

9.  Multicenter laboratory evaluation of the bioMérieux Vitek antimicrobial susceptibility testing system with 11 antimicrobial agents versus members of the family Enterobacteriaceae and Pseudomonas aeruginosa.

Authors:  G V Doern; A B Brueggemann; R Perla; J Daly; D Halkias; R N Jones; M A Saubolle
Journal:  J Clin Microbiol       Date:  1997-08       Impact factor: 5.948

10.  Comparison of sensititre dried microtitration trays with a standard agar method for determination of minimum inhibitory concentrations of antimicrobial agents.

Authors:  D S Reeves; A Holt; M J Bywater; R Wise; M N Logan; J M Andrews; J M Broughall
Journal:  Antimicrob Agents Chemother       Date:  1980-12       Impact factor: 5.191

View more
  7 in total

1.  Evaluation of Sensititre automated reading and incubation system for automated reading of Sensititre broth microdilution susceptibility plates.

Authors:  Kimberle C Chapin; Michael C Musgnug
Journal:  J Clin Microbiol       Date:  2004-02       Impact factor: 5.948

2.  Direct susceptibility testing of positive blood cultures by using Sensititre broth microdilution plates.

Authors:  Kimberle C Chapin; Michael C Musgnug
Journal:  J Clin Microbiol       Date:  2003-10       Impact factor: 5.948

3.  A biosensor platform for rapid antimicrobial susceptibility testing directly from clinical samples.

Authors:  Kathleen E Mach; Ruchika Mohan; Ellen Jo Baron; Mei-Chiung Shih; Vincent Gau; Pak Kin Wong; Joseph C Liao
Journal:  J Urol       Date:  2010-11-12       Impact factor: 7.450

4.  Two-center collaborative evaluation of performance of the BD phoenix automated microbiology system for identification and antimicrobial susceptibility testing of gram-negative bacteria.

Authors:  Maria Grazia Menozzi; Ulrich Eigner; Silvia Covan; Sabina Rossi; Pietro Somenzi; Giuseppe Dettori; Carlo Chezzi; Anne-Marie Fahr
Journal:  J Clin Microbiol       Date:  2006-09-27       Impact factor: 5.948

5.  Evaluation of a scanner-assisted colorimetric MIC method for susceptibility testing of gram-negative fermentative bacteria.

Authors:  Mokhlasur Rahman; Inger Kühn; Motiur Rahman; Barbro Olsson-Liljequist; Roland Möllby
Journal:  Appl Environ Microbiol       Date:  2004-04       Impact factor: 4.792

6.  Antibiotic susceptibility testing of grown blood cultures by combining culture and real-time polymerase chain reaction is rapid and effective.

Authors:  Judith Beuving; Annelies Verbon; Firza A Gronthoud; Ellen E Stobberingh; Petra F G Wolffs
Journal:  PLoS One       Date:  2011-12-14       Impact factor: 3.240

7.  Epigenetic Influence of Dam Methylation on Gene Expression and Attachment in Uropathogenic Escherichia coli.

Authors:  Stacy Ann-Marie Stephenson; Paul D Brown
Journal:  Front Public Health       Date:  2016-06-27
  7 in total